Event
Possible cause
Remedy
The wheelchair cannot be driven.
The battery charger is connected.
Disconnect the battery charger.
The wheelchair is in freewheel mode.
Exit freewheel mode and restart the
wheelchair.
The wheelchair is locked.
Unlock the wheelchair.
A seat function has triggered the
automatic drive inhibit.
Make sure that the backrest and the leg
rest are not tilted and that the seat is not
tilted or lifted.
The motors have gotten too warm.
Turn off the wheelchair and wait for the
motors to cool down.
The wheelchair switches itself off after a
certain period of inactivity (20–30
minutes).
The energy saving mode has been
activated.
Press the start button on the control panel
to turn on the wheelchair again.
The wheelchair stops while being driven.
The main circuit breaker has tripped.
Reset the main circuit breaker once. See
page 43.
The wheelchair can only be driven at
reduced speed.
A seat function has triggered the
automatic speed restriction.
Make sure that the backrest and the leg
rest are not tilted and that the seat is not
tilted or lifted.
72