Section J: Troubleshooting
7
E:
T
ROUBLESHOOTING
Symptom
Check
Action
120 VAC heater jacket is not
heating
Check the resistance between the
black and white wires (tied
together) and the blue wire.
Resistance should be between
100-130 ohms together or from
230-250 ohms for each leg.
If the measure is open for
heater resistance the fusible
link has blown and the jacket
should be replaced.
220 VAC heater jacket is not
heating
Check the resistance between the
black and white power wires using
an ohmmeter. Resistance should
be between 460-480 ohms.
If the measure is open for
heater resistance, the fusible
link has blown and the jacket
should be replaced.
Filter plug cannot be removed
from filter housing
Check “O” rings for damage
High particulate loading
Replace “O” rings
Clean the “O” ring sealing
surfaces with a clean towel
prior to reassembly.
Air from blowback tank keeps
leaking onto sample when not
activated
Voltage going to solenoid is
15VAC or less
Disassemble solenoid valve
and clean flapper, if still leaking
replace valve
Heater clamps not getting hot
Check for 110VAC going into the
Clamps
Replace clamp
For further service assistance, contact: