80 . Spotlight 400 Series User’s Guide
Fitting the ATR Crystal Holder to the Imager
1.
Switch off the imager at the electricity supply.
2.
Stop any laser radiation from entering the imager by switching the internal beam of the
spectrometer to the internal sample compartment.
For further information on how to do this, see the Spectrum software on-screen Help.
3.
Hold the crystal assembly under the upper cassegrain.
Be careful not to change the alignment of the two adjusting levers.
4.
Fit the two adjustment thumb-nuts (Figure 31).
The assembly is aligned if the adjusting levers have not been moved.
5.
Direct the infrared beam to your imager.
6.
Switch on the imager.
ATR cassegrain
ATR crystal holder
Crystal assembly
Knurled
thumb-nut
Figure 31 The ATR crystal holder and ATR cassegrain
Summary of Contents for Spotlight 400 Series
Page 1: ...Spotlight 400 Series User s Guide MOLECULAR SPECTROSCOPY...
Page 6: ......
Page 7: ...Introduction...
Page 13: ...Warnings and Safety Information...
Page 26: ...26 Spotlight 400 Series User s Guide...
Page 27: ...Overview of the Spotlight Imaging System...
Page 39: ...Getting Ready to use the Imaging System...
Page 46: ...46 Spotlight 400 Series User s Guide...
Page 47: ...Preparing Samples...
Page 62: ...62 Spotlight 400 Series User s Guide...
Page 63: ...Techniques for Collecting IR Images and Spectra...
Page 94: ...94 Spotlight 400 Series User s Guide...
Page 95: ...Maintenance...
Page 106: ...106 Spotlight 400 Series User s Guide...
Page 107: ...Appendices...