Routine Maintenance . 65
Changing the Desiccant
CAUTION
Expect to change the desiccant
in the
spectrometer every six months.
Old, used desiccant releases moisture.
In regions experiencing high humidity levels we recommend that you
change the desiccant more often.
The desiccant change interval is set in the
Desiccant change due in (days)
option on the
Setup Instrument BeamPath tab in Spectrum software (Figure 29).
Figure 29 Setup Instrument BeamPath tab in Spectrum
If an instrument desiccant change is overdue, a warning will be displayed in the Status Bar
and in the Setup Instrument BeamPath table.
The warning is displayed until you tell the software that the desiccant has been changed. To
reset the desiccant change interval, select the current value, click
Changed
to clear any
warning messages, and then enter the number of days before the next desiccant change is
due.
Summary of Contents for Frontier IR
Page 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Frontier IR Dual Range Systems User s Guide...
Page 11: ...Warnings and Safety Information...
Page 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Page 34: ...34 Frontier IR Dual Range Systems User s Guide...
Page 35: ...Unpacking and Installation...
Page 45: ...Using the Spectrometer with Spectrum...
Page 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Page 57: ...Routine Maintenance...
Page 78: ...78 Frontier IR Dual Range Systems User s Guide...
Page 79: ...Advanced Maintenance...
Page 94: ...94 Frontier IR Dual Range Systems User s Guide...
Page 95: ...Appendices...