Routine Maintenance . 59
Moving the Spectrometer
WARNING
The spectrometer is a heavy precision instrument, so two people are
required for safe handling.
Consult the local codes of practice issued by safety advisors before
attempting to lift it.
Take care not to injure yourself or others, or to drop the spectrometer.
WARNING
Before moving the spectrometer, switch off the power supply, wait
60 seconds, and disconnect the power cable.
The spectrometer can be lifted using the shaped handholds on its sides, as shown in
Figure 23. Two people are needed to lift it because its basic weight is approximately 36 kg.
Figure 23 Lifting a Frontier IR spectrometer
CAUTION
Do not attempt to lift the instrument, or to move it to another location,
when an external accessory is attached. The optical alignment of the
accessory may be disturbed.
The accessory is fitted to an external accessory bracket mounted within
the handhold on the right of the instrument (Figure 23). The two screws
that attach the external accessory to this bracket are accessible from the
right side of the accessory.
Lifting handhold,
(repeated on left of
the instrument)
Summary of Contents for Frontier IR
Page 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Frontier IR Dual Range Systems User s Guide...
Page 11: ...Warnings and Safety Information...
Page 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Page 34: ...34 Frontier IR Dual Range Systems User s Guide...
Page 35: ...Unpacking and Installation...
Page 45: ...Using the Spectrometer with Spectrum...
Page 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Page 57: ...Routine Maintenance...
Page 78: ...78 Frontier IR Dual Range Systems User s Guide...
Page 79: ...Advanced Maintenance...
Page 94: ...94 Frontier IR Dual Range Systems User s Guide...
Page 95: ...Appendices...