Spectrometer Installation
2
-6
0993-5255
Reinstalling the spectrometer
Reinstall the spectrometer by following exactly the procedures described in this
chapter, as summarized in the steps below:
1. Prepare the new site for the spectrometer; see
Make sure that there will be enough space at the rear and sides of the
spectrometer system for air to circulate freely.
2. Remove the handles. Save the handles and screws.
Insert the plastic plugs provided (B015-3803) into the screw holes.
3. Connect the burner and furnace gas supplies.
4. Connect the drain system to the burner.
5. Install the furnace autosampler.
6. Connect the cooling water supply.
7. Make sure that the main power switches of all the instruments are in the off
position, then connect the electrical cables.