Spectrometer System Description
7
-16
0993-5255
Electrical
protection
As defined in EN 61010-1-1993 (IEC 1010-1):
Insulation: Class I
Insulation category (overvoltage category): II
Pollution degree: 2
Safety standards As defined in EN 61010-1-1993 (IEC 1010-1: 1990 + A1 + A2) and
EN 61010-2-061:1996.
CSA standards: CAN/CSA-C22.2 No. 1010.1-92, Amendment 1 (1992),
2 (1995), and CAN/CSA 61010-2-061 (1995).
UL standards: UL 3101-1 1993 and UL 3111-1 1994.
Compliant to the requirements of Canadian Standards Association (CSA)
and Underwriters Laboratory (UL) as listed in File No. LR 45111.
Certificate of Approval and GS mark from the LGA Bavaria, FRG, Institute
of Equipment Safety Testing and Certification, and CSA-NRTL/C mark
from CSA, Canada, File No. LR 45111.
EMC standards As defined in EN 61326-1:1997.
Environmental
requirements
Ambient temperature: +15 °C to +35 °C
20–80% relative humidity; non-condensing
Dimensions Width: 1095 mm
with
handles:
1220 mm
Height: 655 mm
with door open:
865 mm
Depth: 700 mm
with handles:
770 mm
with autosampler: 1040 mm
Mass
(weight)
147 kg
Spectrometer alone, not including the computer, autosampler, and cooling
system.
Summary of Contents for AAnalyst 700
Page 2: ......
Page 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Page 8: ...Contents C 4 0993 5255...
Page 10: ...1 2 0993 5255...
Page 44: ...Safety Regulatory Information 1 36 0993 5255...
Page 46: ...Spectrometer Installation 2 2 0993 5255...
Page 68: ...Spectrometer Installation 2 24 0993 5255...
Page 70: ...Preparing the System 3 2 0993 5255...
Page 84: ...System Maintenance 4 2 0993 5255...
Page 98: ...Parts and Supplies 5 2 0993 5255...
Page 102: ...Laboratory Requirements 6 2 0993 5255...
Page 120: ...Spectrometer System Description 7 2 0993 5255...
Page 136: ...Translations of Warnings 8 2 0993 5255...
Page 146: ...Translations of Warnings 8 12 0993 5255...
Page 185: ...0993 5255 In 1 Index...
Page 186: ...Index In 2 0993 5255...
Page 192: ......
Page 193: ......