XRpad2 4336
14
68153 01
D
IGIT
A
L
I
MAGING
includes the specification for the XRpad2 LBP.
8.5.1.1
Charging the XRpad2 LBP
•
A new rechargeable Lithium Battery Pack (XRpad2 LBP) comes in a discharged condition
and must be charged using the dedicated lithium battery charger (XRpad LBC) before use.
Refer to the XRpad LBC manual for more details.
•
The XRpad LBC will charge the XRpad2 LBP to a usable condition within three hours
depending on the initial state of charge. The X-ray a detector, when connected to the XRpad
Interface and Power Unit (XRpad IPU-2), can also charge the XRpad2 LBP, but the charge
rate is much slower.
•
A charged battery will eventually lose its charge if unused. Upon initial use (or after a
prolonged storage period), the battery may require three to four charge/discharge cycles
before achieving maximum capacity.
•
The actual battery run-time will depend on the power demands made by the X-ray detector.
•
The XRpad2 LBP is keyed and can only be inserted into the XRpad LBC charger in one
orientation.
Table 9
Specification of the XRpad2 LBP
General
REF
95510949H-01
Voltage
11.1V
Amp-hours
4.2Ah
Capacity
46.6Wh
Charging Time
Within 3 hours
Environmental
Relative Humidity
35% to ~85%
Operating (discharging)
-10°C to 55°C
Charging
0°C to 42°C
Transportation
-20°C to 45°C
Storage (up to 3 months)
15°C to 35°C
Ingress
Ingress Protection Rating
IP54
Lifetime
Charge-discharge Cycles
500 cycles on average under normal usage conditions.
Battery should be discarded on or before five years from
date of manufacture indicated on the battery label.