6. TEMPERATURE METER
CALIBRATION SETUP
6.2 0 °C Calibration
6-2-1.
Press and hold “Power” + “Light” buttons for 2
seconds to enter setup mode.
6-2-2.
Press “REL” button to enter calibration mode.
Enter PT selection first.
6-2-3.
Press “°C/°F” button to confirm PT selection.
6-2-4.
Press “REL” button to enter calibration mode.
Insert temperature probe into 0°C Deg.
Standard calibration unit before
calibration. Hold the probe until condition
is stabilized before starting calibration.