6. BUTTON INSTRUCTIONS 

P.8

 

6.1 Power ON/OFF and Back-Light Button: 

Press the       button to turn on the meter. Press and hold the button for 3 

seconds to turn off. 
Press       to turn on the LCD backlight. This makes it easier to read in dark 

environment. Press it again to turn off back-light. The back-light will be 

automatically turned off after 30 seconds to save battery power.

6.2 Data-Hold Button :

Press        button to freeze the data shown on the LCD screen. Press it 

again to exit Data-Hold mode.

Note: 

 When the unit is in the Data-Hold mode,        ,        , and        buttons 

are disabled.

6.3 REC Button:

Press        button to start recording data. The 

REC

 symbol will display on 

the screen. To stop recording, press        button again.

Note: 

 During recording period, most of the buttons are disabled, such 

as the        . All settings must be made before starting the data 

logger function.

Note: 

 When the memory is full (32,000 recorders), 

FULL

 symbol will 

blink on the LCD screen. The data logger stops.

Note: 

 When battery power is low ( “          ” symbol lights up on the 

screen), data logger cannot be started. If the battery is running 

low during data logging, it will stop recording automatically.

 

Summary of Contents for CO2512

Page 1: ...Datalogger Carbon Dioxide Meter CO2512 Instruction Manual...

Page 2: ...s 7 2 Menu Item 7 3 Menu Description 7 3 1 Setting interval time for data storing 7 3 2 Set offset to compensate for errors 7 3 3 Set altitude compensation 7 3 4 Set alarm point 7 3 5 Set auto power o...

Page 3: ...efully before attempting to operate or service the meter Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired ENVIRONMENT CONDITIONS Altitude u...

Page 4: ...CO2 sensor Non Dispersive Infrared for superior stability User operated CO2 calibration Data Hold Max Min STEL TWA function Adjustable auto power off timer Internal CO2 alarm buzzer Datalogging up to...

Page 5: ...Sampling rate Non dispersive infra red NDIR 0 30000 ppm 75ppm 3 of reading 0 5000 ppm 150ppm 5 of reading 5001 30000 ppm 1 ppm 20 s 2 s CO2 SENSOR Measurement range Accuracy Resolution Response time...

Page 6: ...line battery Approx 10 hours alkaline battery 0 C to 50 C 32 F to 122 F 10 to 90 RH no condensing 20 C to 60 C 4 F to 140 F 10 to 75 RH 226 5 L x65 W x38 5 H mm Approx 273g 9V alkaline battery Instruc...

Page 7: ...n indicator Auto power off enabled indicator Recording data logger indicator Memory full indicator CO2 reading Relative Humidity Temperature reading CO2 alarm indicator Hi CO2 alarm indicator Lo CO2 a...

Page 8: ...splay Screen Power and Back Light Button MAX MIN AVG Button SETUP Button DATA HOLD Button 2 3 4 5 6 1 REC Button C F Button Calibration inlet USB Interface Tilt Stand Battery Compartment 8 9 10 11 12...

Page 9: ...n to exit Data Hold mode Note When the unit is in the Data Hold mode and buttons are disabled 6 3 REC Button Press button to start recording data The REC symbol will display on the screen To stop reco...

Page 10: ...on LCD 3 Press button again to show average data the STEL symbol lights up on LCD 4 Press button again to show average data the TWA symbol lights up on LCD 5 Press button again the MAX MIN STEL and TW...

Page 11: ...ETUP Button Press the button to enter setup options Press again to exit setup mode at anytime Note When the meter is connecting with PC setup options cannot be entered 6 6 C F Button Press to select t...

Page 12: ...ms 3 Press the button to save changes and move to next setting option 7 2 Menu Item Note Do not hold the meter close to faces in order to avoid exhalation affects reading CO2 1 Fig 1 Set interval time...

Page 13: ...ute value 2 Press the button 3 Press the button to increase decrease second value 4 Press the button to set interval Note Set interval time range 1 Second 00 01 to 60 minutes 59 seconds 60 59 7 3 2 Se...

Page 14: ...ase 4th digit value 0 9 8 Press the button to set altitude compensation 7 OPERATING INSTRUCTIONS P 13 3 Press the button to increase decrease temperature value 4 Press the button 5 Press the button to...

Page 15: ...Fig 11 Fig 13 Fig 14 Fig 12 Note When the measuring values go beyond alarm point the symbol will blink Hi or Hi on the LCD The beeper will make beep beep beep sound The Lo alarm set cannot be greater...

Page 16: ...the clock setting after replacing batteries If necessary reset system clock 7 4 Clearing Data Logger Records 1 Turn off the unit 2 Press and hold button and then press power button to turn on the unit...

Page 17: ...the range of each sensor Do not use gas that could over range the sensor For better accuracy it is recommended that you use a span gas value close to the gas concentration you expect to measure Please...

Page 18: ...onds press to enter calibration mode 5 LCD line1 indicator CO2 real time reading 6 LCD line2 symbol blinking Press the button to increase decrease value 7 Please wait at least 2 minutes and then press...

Page 19: ...range the symbol will blink on the LCD display and the battery needs to be replaced 2 Before replacing the battery power off the meter Open the cover of the battery cabinet Replace the old batteries...

Page 20: ...a clean lint free antistatic and dry cleaning cloth Do no use cleaning agents that contain carbon or benzenes alcohol or anything similar to clean the product since these substances damage the surfac...

Page 21: ...se SE512 Recording real time data in waveform 1 Power on the NDIR CO2 Meter first and connect it to a PC USB port with the cable 2 Start SE512 program 3 If the connection is successful the panel will...

Page 22: ...g window for you to choose the file name and file type to save There are three types of file format you can choose binary file ghf text file txt and EXCEL format file csv The ghf file use much fewer d...

Page 23: ...file 1 Power on the Carbon Dioxide CO2 Meter 2 Press the REC button of the meter to start recording data 3 After a while press REC button again to stop recording data 4 Connect the meter to PC 5 Star...

Page 24: ...Panel Window the user can control meter via the button in this window Open Real Time Graph display to graph the present data Start collecting real time data Stop collecting real time data By opening...

Page 25: ...tool bar to load recorded data from the meter and there will be a progress indicator to show the loading status If error occurs just click DataLogger icon again After the data was downloaded the left...

Page 26: ...also click at any other data set for graph Graph Tool Bar Display or hide Statistic1 Display or hide Statistic2 Normal cursor When selected the mouse cursor will become a cross sign when moving to the...

Page 27: ...channel Graph Customization Change the Y axis extention To Undo the Zoom To Zoom 1 Press the left mouse button and drag the cursor to select the new extents 2 Release the mouse button You can choose c...

Page 28: ...2 How to zoom the graph Answer Press the left mouse button and drag the cursor to select the new extents then release the mouse button 3 When I setup the real time sampling with a fast rate eg 1 sec s...

Page 29: ...espread 1000 ppm should be used as an upper limit for indoor levels ASHRAE Standard 62 1989 1000ppm CO2 concentration in occupied building should not exceed 1000ppm Building bulletin 101 BB101 1500ppm...

Page 30: ...r products and services please send us an email cs perfectprime com For B2B or project based application please send an email sales perfectprime com FOR MORE INFORMATION ABOUT PERFECTPRIME PLEASE VISI...

Page 31: ...HORMES LIMITED cs perfectprime com G F UNIT 3 61 GLENTHORNE ROAD LONDON W6 0LJ UNITED KINGDOM 44 203 7695377 Retailer Email Address Telephone...

Reviews: