For additional information visit
www.pepetools.com
6
Program setup
Configure program
Programs 1 to 5.
Line 1 – Select current active program
Line 2 – Reverse time from 0 – 90 minute, step 1 sec
Line 3 – Plate time from 0 to 9 hours and 59 minute, step 1 sec
Line 4 – Temperature from 0-99 °C (0 – 210°F depending of temperature units setting), when is 0, device will not turn on the
heater*)
Line 5 – Plate voltage – from 0.5 to 12V (for P12V25Ax)
Line 6 – Plate current limit – from 0 to 25 A (for P12V25Ax)
Line 7 – Delay after end of plate process
Line 8 – Exit - Back to main menu
Note: Depending on configuration, when some of features are disabled or missing (for example out-
put voltage reverse) menus for related option will not display
.
*
Available if the device has a Heater option.
Setting Sequence program
Configure program from Sequence (programs 1 - 5 to execute sequentially).
Line 1 – Select current active program.
Line 2-6 – Step 1-5 - select sequence of the programs.
Line 7 – Exit - Back to main menu.
Manual mode
Manual control of plating, heating, output voltage, current limit and output voltage reverse
Line 1 – Manual mode and polarity symbol indication
Line 2 – Output voltage and polarity
Line 3 – Output current and plate ON/OFF state
Line 4 – Exit – Back to main menu
Note:
Depending
on configuration, when some of features are disabled, missing or not used (for example heating), the menus
for related option will not display.
The symbol “ “ show current selected item.