2-3. PREPARATION JUST BEFORE INSERTION OF ENDOSCOPE
1) If the endoscope has just recently been reprocessed, has been prepared or stored properly and passed all pre-procedure
inspections, the instrument should be ready to use. If necessary, the scope’s insertion tube may be wiped down with a gauze
dampened with 70-90% ethyl or isopropyl alcohol.
2) Gently clean the objective lens with a cotton-tip applicator moistened with 70-90% ethyl or isopropyl alcohol. A lens cleaner
(anti-fogging agent) may also be applied via gauze or other applicator.
3) Check the endoscopic image and confirm that it is of acceptable quality for clinical use. Refer also to the owner’s manual
supplied with the PENTAX video processor for inspection of the image quality.
4) Prior to trans-oral insertion of the endoscope, place a bite-block (mouthpiece) into the patient’s mouth to protect the endoscope
from damage during the procedure. Failure to do so can result in scratches, tears and/or crushing of the insertion portion of the
endoscope.
5) Apply a medical grade water soluble lubricant to the insertion tube. Do not use petroleum based lubricants.
WARNING:
Every endo
s
cope
s
ho
u
ld be properly di
s
infected or
s
terilized before being
us
ed for the fir
s
t time. The endo
s
cope
s
ho
u
ld h
a
ve been properly cle
a
ned
a
nd di
s
infected or
s
terilized
a
fter
a
ny previo
us
us
e
a
nd
a
fter being ret
u
rned
for
a
ny rep
a
ir
s
/
s
ervice.
Refer to the comp
a
nion In
s
tr
u
ction
s
for U
s
e (Reproce
ss
ing) de
s
cribing in det
a
il PENTAX reproce
ss
ing
in
s
tr
u
ction
s
.
WARNING:
C
u
rrent infection control g
u
ideline
s
req
u
ire th
a
t endo
s
cope
s
a
nd their p
a
tient cont
a
ct
a
cce
ss
orie
s
either be
s
terilized or
a
t the le
as
t be
su
bjected to high-level di
s
infection. Acce
ss
orie
s
which ENTER
S
TERILE TI
SS
UE or
THE VA
S
CULAR
S
Y
S
TEM m
us
t be
s
terilized before p
a
tient
us
e. Only the
us
er c
a
n determine if
a
ny in
s
tr
u
ment
s
a
nd
a
cce
ss
orie
s
h
a
ve
u
ndergone
a
ppropri
a
te infection control proced
u
re
s
prior to e
a
ch clinic
a
l
us
e.
NOTE:
Cont
a
ct the m
a
n
u
f
a
ct
u
rer
a
nd follow loc
a
l reg
u
l
a
tion
s
reg
a
rding
sa
fe
us
e,
a
ppropri
a
te h
a
ndling
a
nd di
s
po
sa
l of
a
lcohol prod
u
ct
s
. M
a
teri
a
l
Sa
fety D
a
t
a
S
heet
s
(He
a
lth
a
nd
Sa
fety D
a
t
a
S
heet
s
or
s
imil
a
r doc
u
ment
s
depending
u
pon co
u
ntry)
a
v
a
il
a
ble from the
a
lcohol m
a
n
u
f
a
ct
u
rer
s
ho
u
ld provide g
u
id
a
nce to end
us
er
s
a
bo
u
t compo
s
ition,
h
a
z
a
rd
s
, chemic
a
l
a
nd phy
s
ic
a
l propertie
s
, fir
s
t
a
id, h
a
ndling
a
nd
s
tor
a
ge,
s
t
a
bility, prec
au
tion
s
, di
s
po
sa
l, etc.
ass
oci
a
ted with
a
lcohol
s
ol
u
tion.
NOTE:
The objective len
s
m
us
t be kept free of the l
u
bric
a
nt
a
nd exce
ss
len
s
cle
a
ner.
WARNING:
Never drop thi
s
eq
u
ipment or
su
bject it to
s
evere imp
a
ct
as
it co
u
ld compromi
s
e the f
u
nction
a
lity
a
nd/or
sa
fety
of the
u
nit.
S
ho
u
ld thi
s
eq
u
ipment to be mi
s
h
a
ndled or dropped, do not
us
e it. Ret
u
rn it to
a
n
au
thorized PENTAX
s
ervice f
a
cility for in
s
pection or rep
a
ir.
Summary of Contents for ED-3490TK
Page 27: ......