8.2 Repair
8.2.1 Replacing the pressure gauge
Method:
1. Unscrew the pressure gauge’s (pos. 67) cover.
2. Using a pressure gauge puller (included in the service set), pull the pressure gauge out
of the instrument whilst rotating it.
3. Take the pressure gauge out of the housing.
4. Check the O-rings (pos. 5 and 36). If necessary, replace them.
5. Carefully install the new pressure gauge.
6. Replace the pressure gauge’s cover.
8.2.2 Replacing the battery
An electrical impulse is given via the platinum wires when the handle (pos. 22) is moved from
the
sample taking
position to the
measuring
position.
This impulse lasts for approximately 7 seconds. If the LED starts to flash, then the battery must
be replaced.
Attention!
There must not be any condensation in the battery housing.
Therefore, first allow the temperature of the instrument to come up
to room temperature.
Method:
1. Unscrew the battery housing’s (pos. 76) cover,
2. Remove the battery from the battery housing,
3. Insert a new battery into the correct position in the battery housing (pay attention to the
+ve and -ve terminals),
4. Replace the battery housing’s cover.
Summary of Contents for HAFFMANS GMT
Page 2: ......
Page 4: ......
Page 6: ...HAFFMANS GMT Page 6 of 37 GMT...
Page 8: ...HAFFMANS GMT Page 8 of 37 GMT...
Page 10: ...HAFFMANS GMT Page 10 of 37 GMT...
Page 12: ...HAFFMANS GMT Page 12 of 37 GMT...
Page 14: ...HAFFMANS GMT Page 14 of 37 GMT...
Page 24: ...HAFFMANS GMT Page 24 of 37 GMT...
Page 26: ...HAFFMANS GMT Page 26 of 37 GMT...
Page 28: ...HAFFMANS GMT Page 28 of 37 GMT...
Page 30: ...HAFFMANS GMT Page 30 of 37 GMT...
Page 32: ...HAFFMANS GMT Page 32 of 37 GMT...
Page 36: ...HAFFMANS GMT Page 36 of 37 GMT...
Page 38: ......