Operating the Anaesthesia System
6.13 System prompt message and icons field
NOTE
Prompt messages are not alarm messages.
Clinical and technical alarm messages and system status
messages are also displayed in the prompt message area.
6.13.1 Patient type icon
The display (1) shows an icon relevant to the selected patient
type.
6.13.2 Power status display field
The power status display indicates mains power (2) and battery
power (3) .
6.13.3 System time field
The time field (4) displays system time data.
6.14 System start-up procedure
1. Plug the power cord into an AC mains power outlet.
The mains indicator is lit when the AC power is connected.
The battery will be charged (if it is not already fully
charged).
2. Check that the breathing system is properly connected.
3. Turn the Anaesthesia system switch (1) to the ON position.
4. The self test will start and the indicator lights will be
illuminated.
The self-test screen will be displayed and the self test will
start.
5. The self-test will continue for about 10 seconds. The
device will prompt the user to perform a pre-use test.
Press the ACCEPT button to access the pre-use test
screen.
1
3
4
2
1