Technical Note 

Technical Note:  PendoTECH Process Control Systems: Pump Calibration 
Revision 0 
Page 1 of 5 

 

 

 

 

PendoTECH Process Control Systems: Pump 
Calibration  

  

1. Introduction 

The PendoTECH Filter Screening System (NFF) and TFF Process (TFF) allow users to automatically 
control flowrates throughout the duration of an experiment or process. Any pump capable of accepting a 
remote-control signal can be interfaced with PendoTECH control systems. PendoTECH systems control 
pumps by sending an analog signal which is proportional to the flowrate entered by the user on the 
system’s graphical user interface (GUI). The analog output is scaled linearly from the pumps minimum 
flowrate to its calculated maximum flowrate. The following note describes the process of configuring a 
pump with its nominal flow values and making any necessary adjustments to compensate for the inherent 
inefficiencies associated with pump technology.  
 

2. Connecting Pumps  

a. PendoTECH NFF System 

The  PendoTECH  NFF 

system  is  capable  of 

independently controlling four pumps simultaneously. A 
custom cable is included with purchase of the system that 
must be specified to match the manufacturer of the pump 
to  be  used  with  the  system,  i.e.  Masterflex  or  Watson-
Marlow. The four pumps must also be of identical make 
and model in order to be used simultaneously. The use of 
an adapter cable is available, upon request, if there is a 
need  to  use  a  pump  of  different  make  and  model.  An 
example of a Masterflex pump cable is shown at right and 
the pump cable input is called out in the figure below. 
 

 

           Figure 2.1 Back Panel of NFF Control System 

 

b. PendoTECH TFF System 

The PendoTECH TFF system can control up to three pumps simultaneously, specifically a main 
circulation pump, a diafiltration pump, and a Filtrate/Permeate pump. Unlike the NFF system, the 
pumps being used on the TFF system do not need to be of the same make and model. However, it 
is important to note that the connection cable between the control system and pump will vary 
between different model pumps The pump inputs for the TFF system are called out below. 

Note - Only one of the following inputs will be used: Circulation Pump/Circ Pump Alternate 