![Pelton & Crane Magna-Clave Use & Care Manual Download Page 18](http://html.mh-extra.com/html/pelton-and-crane/magna-clave/magna-clave_use-and-care-manual_1523574018.webp)
18
1.
PREVENTIVE MAINTENANCE SCHEDULE
CHECK
FREQUENCY
PROCEDURE
ACTION
Clean and inspect
Weekly
12 oz.. Omni-Cleaner
See Sections 4-Il and
chamber
to I gallon distilled water
4-IV. Call authorized
on 20 minute cycle.
service representative
cracks or fissures are
found. Do not operate
unit.
Safety valve ring
Every 3 months
Remove casing plate (located
If valve does not
WARNING: When
on top rear of unit). Manually
open, turn off POWER
ring is pulled on
pull ring while chamber is un-
and call for service
safety valve with
der pressure.
representative.
unit under pressure
steam is discharged
from the chamber
at a high tempera-
ture. Keep clear of
steam discharge
path to avoid burns.
Use of a hot pad or
instrument to pull
ring is suggested.
Door gasket
Weekly
Inspect and clean using
Call authorized service
Omni-Cleaner or mild
representative for re-
detergent and distilled water.
placement gaskets.
Check for leaks (have
leaking gasket replaced).
Chamber fill filter
Weekly
Inspect and clean.
See Section 4-Il.
Door interlock
Weekly
Inspect. See Section 4-II.
Call authorized service
representative for im-
proper closure or signs
of wear.
Boiler ring
Weekly
Inspect and clean using
nonchlorinated pad which
contains no metal.
SECTION 3