2
Transport and Storage Environmental Conditions (not to exceed 15 days)
Ambient temperature range of -40°F (-40°C) to +158°F (+70°C)
Relative humidity range of 10% to 100%, including condensation
Atmospheric pressure range of 50KPa to 106KPa
The Pedigo Warming Cabinet has been
thoroughly tested, checked for calibration,
and inspected to insure only the highest
quality. When you receive your unit, check
for any possible shipping damage and
report it at once to the delivering carrier.
The cabinet, complete with unattached items and
accessories, may be delivered in one or more packages. Save all
the information and instructions packed inside the cabinet.
Complete and return the warranty card to the factory as
soon as possible to assure prompt service in the event of a
warranty parts and labor claim.
All claims for warranty must include
the full model number and serial number of the cabinet.
After unpacking the unit, allow it to remain under stable
atmospheric conditions for several hours before connecting to
the power source. Extreme changes in temperatures and/or
humidity during storage and/or transportation could have
allowed condensation to form in the unit.
NOTE
:
Pay particular attention throughout this manual to
the warning and caution symbol which
means: Attention, Consult Operation
Manual before proceeding.
E L E C T R I C A L I N F O R M AT I O N
The power specifications are located on the unit
identification nameplate. This nameplete is permanently
attached to the unit and must be located to determine and
verify power requirements.
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
1000 Watts, 8.0 Amps
The single compartment warming cabinet is constructed with 22
gauge stainless steel exterior casing and door, with handle and
hinges designed to withstand heavy usage. The cabinet is heated
by the Halo Heat system incorporating low watt density cable
mounted around the compartment walls. This system provides
evenly distributed heat without the need of a heat-circulating fan.
The cabinet is controlled by an adjustable, electronic thermostat,
which includes a heat indicator L.E.D., temperature set button,
temperature increase and decrease button, and digital display to
monitor compartment temperature. The thermostat has a
temperature range of 90°F (32°C) minimum to 200°F (93°C)
maximum. The chamber is equipped with one (1) white, epoxy-
coated blanket support assembly with shelf, and furnished with
one (1) set of 5" (127mm) heavy-duty casters — two rigid and two
swivel with brake.
P R E PA R AT I O N
Before operating the cabinet, clean both the interior and exterior
of the unit with a damp cloth and mild soap solution. Wipe with
an appropriate disinfectant. Clean and install the cabinet shelf
assembly. Do not install or operate the blanket warmer in close
proximity of flammable anesthetics.
B L A N K E T WA R M I N G C A B I N E T
T R A N S P O R T a n d S T O R A G E
G E N E R A L I N F O R M AT I O N
N E M A 5 - 1 5 P
15A - 125V Plug
Hospital Grade
Grounding reliability can only be achieved when equipment is
connected to an equivalent receptacle marked "Hospital Grade."
Protective Earth
Ground Symbol
ENSURE POWER SOURCE
MATCHES VOLTAGE STAMPED
ON UNIT NAMEPLATE.
Hazardous
Voltage Present
At no time should the inside or outside of
the cabinet be washed down by flooding with
water or liquid solution.
N E V E R S T E A M
C L E A N .
S e v e r e d a m a g e o r e l e c t r i c a l
hazard could result.
Explosion risk if used in
the presence of
flammable anesthetics.
Danger!
U N P A C K I N G A N D S E T - U P
Medical Equipment classified by Underwriters
Laboratories with Respect to Electrical Shock, Fire
and Mechanical Hazards only, in Accordance with
UL 544 and CAN/CSA C22.2 No. 601.1.
IP50
MADE IN USA
MODEL
SERIAL#
WATTS
AMPS
VOLTS
1 PH
Hz
LISTED
MEDICAL EQUIPMENT
C
LISTED
US
MEDICAL EQUIPMENT
1P50
Pedigo P-2030-S Blanket Warming Cabinet Operation and Care Manual