IEEE488.2 Common Commands
13
[SENSe:]VOLTage[:DC]:IMPedance:AUTO
Syntax
[SENSe:]VOLTage[:DC]:IMPedance:AUTO {OFF|ON}
[SENSe:]VOLTage[:DC]:IMPedance:AUTO?
Description
Close or start auto input impedance mode for DC measurement
Parameter
Name
Type
Range
Default
<bool>
bool
{OFF|ON}
OFF
Note
OFF: for all the measuring range, DC voltage input impedance
is fixed at 10 MΩ
ON: DC voltage measurement input impedance changes in accordance with measuring range.
Input impedance is 10
GΩ at 200mV and 2V measuring range, 10 MΩ at 20V, 200V and 1000V
measuring range.
Return format
Return 0 (OFF) or 1 (ON) after query
[SENSe:]CONT:THREshold
Syntax
[SENSe:]CONT:THREshold <values>