4. Measurement Operation
1. Insert the microphone of your sound level meter in the
calibration port of the PeakTech 8010. Make sure that the
microphone is fully inserted into the calibration port.
2. Select the required sound level for testing (94 or 114dB).
3. Now read the measured value on the display of your sound
level meter from.
4. If the measured value does not match with the calibration
signal, you can readjust the sound level meter. For this
purpose refer to the manual of your sound level meter or ask
your dealer for it.
5. After every measurement with the sound level calibrator, it
should always be turned off, by choosing the "OFF" position
of the selector switch.
5. Technical specifications
Output signal
94 dB or 114 dB
sound pressure level
Output frequency
1000 Hz; +/-4%
Reference conditions
Temperature: 23°C (73°F)
Pressure: 1013 mbar
Humidity: 65% RH
Accuracy
+/- 0,5 dB
Low Battery test
The calibrator stops the output of
a sound pressure level, when the
battery voltage falls below the
minimum requirement.
This instrument complies with
the standards
EN 60942:2003 Class 2
CE:
EN 61326-1, EN 61326-2-1
Dimensions (WxHxD)
51 x 120 x 43mm
Weight
130 g
-11-