description. If it is found that there is any accessory lost or damaged, please get in
touch with your distributor.
3. Check the Complete Instrument
If it is found that there is damage to the appearance of the instrument, or the
instrument cannot work normally, or fails in the performance test, please get in touch
with your distributor. If there is damage to the instrument caused by the transportation,
please keep the package. With the transportation department or distributor
responsible for this business informed about it, a repairing or replacement of the
instrument will be arranged.
4.5. Foot Stool Adjustment
Unfold the foot stools on the bottom of the generator, as page 60 Figure 0-1 - 11.
4.6. AC Power Input Setting
The waveform generator adopt 110V/220V AC power source. Users should regulate the
voltage scale of the
Power Switch
according to the standards in their own country (see
Figure 0-2 / 8 on page 61) at the rear panel and use an appropriate fuse.
Voltage
Fuse
110V
125 V, F4AL
220V
250 V, F2AL
To change the voltage scale of the instrument, do the following steps:
1. Turn off the power button at the front panel and remove the power cord.
2. Check if the fuse installed before leaving factory (250 V, F2AL) can match with
the selected voltage scale; if not, pry the cover open using a straight screwdriver
(Figure 0-2 / 7 on page 61) change the fuse.
3. Regulate the
Power Switch
to the desired voltage scale.
4.7. Power On
(1) Connect the instrument to the AC supply using the supplied power cord.
Warning:
To avoid electric shock, the instrument must be grounded properly.
(2) Press down the
power button
at the front panel, the screen shows the boot
screen.