3. Remove the old fuse from its holder by gently pulling it out.
4. Install the new fuse into the holder.
5. Always use a fuse of the proper size and value
mA-range: 500mA/690V (10,3 x 38mm) fast blow
10A-range: 10 A/690V (10,3 x 38mm) fast blow
6. Put the back case back in place. Insert the screws and
tighten it securely.
Warning:
To avoid electric shock, do not operate your meter until the back
cover and the back case is in place and fastened securely.
6.3. Care
*
Get in the habit of double checking the position of the switch
before making a measurement. The meter can be burned
out by applying voltage when the switch is set on the current
or ohms range.
*
If the unit has not been in use for a long period of time,
rotating the switch in both direction several times will wipe
the contacts for a good and clean connection.