40
3.3. DC Current
Range
Resolution Load Voltage
Accuracy
2 mA
0.001mA
104mV/µA
± 0,8% + 3 dgt.
20 mA
0.01mA
12.3mV/µA
200 mA
0.1mA
3.97mV/µA
± 1,2% + 4 dgt.
10 A
0.01A
265mV/A
± 2,0% + 5 dgt.
Overload protection:
0,2A / 1000V: 6,3 x 32 mm fuse in mA-Input
10A / 1000V: 10,3 x 38 mm fuse in 10A-Input
10A for max. 10 sec. every 15 min.
3.4. AC Current
Range
Resolution
Accuracy
2 mA
1 µA
± 1,0% rdg. + 5 dgt.
20 mA
10 µA
200 mA
100 µA
± 2,0% rdg. + 5 dgt.
10 A
10 mA
± 3,0% rdg. + 10 dgt.
Overload protection
0,2A / 1000V: 6,3 x 32 mm fuse in mA-Input
10A / 1000V: 10,3 x 38 mm fuse in 10A-Input
10A for max. 10 sec. every 15 min.
Frequency - range: 40 ... 200Hz
Crest factor: CF≤3, when CF≥2, add an additional error of
1% of the reading