NM60L NITROGEN GAS GENERATOR

 

 

 

 

 

 

USER MANUAL 

UM-NM60L -19”Rack

 

Summary of Contents for NM60L

Page 1: ...NM60L NITROGEN GAS GENERATOR USER MANUAL UM NM60L 19 Rack ...

Page 2: ...y Page No 3 Warranty Statement 4 Safety Notice to Users 4 Declaration of Conformity 5 Air Connection 6 Principle of Operation 7 Commissioning 8 Maintenance Schedule 9 Technical Specifications 12 Pneumatic Diagram 13 Over Dimension 14 Maintenance Log 15 Notes 16 2 ...

Page 3: ...ls Date 1 Document Created H D 30 05 03 2 USA Technical support number updated FAD 10 11 04 3 Document Updated FAD 11 03 05 4 New Style Front Added FAD 08 04 05 5 Added Overall Dimension S Khan 05 06 08 6 Updated Diagram S Khan 05 05 10 7 Filter Info Update LC 01 05 13 3 ...

Page 4: ...the Company 5 Subject to the Company being notified of any defect as is referred to in sub clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub clause 4 hereof the Company shall in its option replace or repair the defective Goods or refund a proportionate part of the Price The Company shall have no further liability to the Customer save as mention...

Page 5: ...contacting us through info peakscientific com WEEE Statement of Compliance WEE FjQ I1 16 X1U The WEEE Waste of Electrical Electronic Equipment Directive 2002f96 EC issued by the European Union aims to reduce the impact upon the envirronment from disposal of certain types of equipment It requires producers to in plement controls to ensure that eq uipment that they produce is correctly disposed of f...

Page 6: ...ogen Outlet connections are at the rear of the generator in the middle right side of the panel The Air inlet connection is the left port 3 8 BSP female outlet port is 1 4 BSP female The generator is supplied with two Breathing Air pre filters and an Activated Carbon post filter which are fitted internally The Breathing Air filters will drain moisture and are equipped with automatic drains This is ...

Page 7: ...ermeate through the membrane wall whilst so called Slow gases will not and continue along the membrane tube and are thus available for collection and use Air passes into the system via the Breathing Air Filter This will prevent bulk moisture and harmful contaminants from reaching the Membrane The clean air then passes into the Membranes where Oxygen CO2 and Moisture are removed allowing High Purit...

Page 8: ... and pressure Once the Membrane reaches the design pressure the Generator will stabilise and produce pure Nitrogen Maximum purity will be achieved after around 1 hour After this time the generator can be re connected to the application The design of the generator is that it will deliver up to rated output flow of Nitrogen at 100 psig Should the demand for Nitrogen be less than the rated output flo...

Page 9: ...ipment it should be regularly inspected and serviced as below BA Inlet Filters Every 12 months RAC Filter Elements Every 12 months BA Inlet Filter Elements These should be changed at 12 months intervals In addition filter bowls should be cleaned and the operation of the auto drains should be checked if auto drain requires replacing Part is 02 4485 FAILURE TO FOLLOW THE PRESCRIBED MAINTENANCE PLAN ...

Page 10: ...nerator MUST be turned OFF and the Generator MUST be de pressurised prior to attempting to remove ANY filter bowl Failure to do this may cause injury Inlet Breathing Air Filter Separator Elements The top Filter Element 00 4427 should be changed at 12 month intervals The bottom element should be changed at 12 month intervals The Part Numbers are as shown in the drawings below RAC Element Prt 00 442...

Page 11: ...turned OFF and the Generator MUST be de pressurised prior to attempting to remove ANY filter bowl Failure to do this may cause injury R A C Elem ent P rt 00 4425 The filter housing is removed by un screwing in a counter clockwise direction IMPORTANT Ensure that the system is NOT under pressure before attempting to remove the bowl and that the air supply is shut off 11 ...

Page 12: ...m Air Inlet Flow Rate 180 l min ATP Outlet Gas Maximum Pressure Drop Outlet Inlet δP 8 psig Maximum Gas Outlet Pressure 100 psig Maximum Outlet Flow High Purity Nitrogen 60 Litres min ATP Start up time for Purity 60 minutes Particles 0 01um Serviceable Parts List Service Kit Part 08 4429 Breathing Air Inlet Filter Elements 00 4424 00 4427 Active Carbon Filter Element 00 4425 General Dimensions W x...

Page 13: ... 0 I J c 3 II A l dM Switch Safety Valve Nort ReturnValve Membranes Inlet Valve Receiver Inlet BA Filter Pressure Regulator Flow Controller RAC Filter w NM 60L System Diagram ...

Page 14: ... II NM60L Nitrogen Gas Generator Overall Dimensions 23 38 inch 594 0 mm c Overall Dimensions cr II II Ill E ISOMETRIC VIEW 14 ...

Page 15: ...NM120L Nitrogen Gas Generator UM NM60L Maintenance log Maintenance Log Model NM60L Serial number Work Done Remarks Date Name 15 ...

Page 16: ...Notes 16 ...