NM20ZL User Guide Issue 2

 

Air Connection

 

 

The Nitrogen Generator should be connected to a 

clean, dry, OIL - FREE

 source of compressed air. A 

regulated minimum pressure of 120 psig is required. Any doubts as to the suitability of your compressed 

air supply should be referred to the factory for advice. 

 

The  generator  has  a 

Breathing  Air  Filter

  with  ¼”  BSPT  connection  to  the  left  side  of  the  unit.  The 

Compressed Air supply should be connected here. This filter will drain moisture and is equipped with an 

automatic  drain.  The  drain  should  be  led  to  a  convenient  place.  There  are  3-off  ¼”  BSPT  Female 

bulkhead  connections  to  the  right  of  the  unit.  Fittings  are  provided  to  connect  these  ports  to  the  gas 

connections of the Mass Spectrometer.  

 

A kit of fittings for connecting to the Mass Spectrometer is provided with this User Guide. 

 

LC/MS with 3-off Connections (e.g. Api2000) 

 

Connect  the  3-off  gas  ports  at  the  right  side  of  the  generator  to  their  respective  ports  on  the  Mass 

Spectrometer using the 3-off ¼” BSPT x ¼” Swagelok fittings provided. Use ¼” tubing throughout. 

 

LC/MS with 4-off Connections (e.g. Api3000)

 

 

Connect 

Curtain Gas

 and 

Exhaust 

to their respective ports using ¼” tubing as per above. The 

Source 

Gas 

will  have  to  be  split  into 

Gases  1  &  2

  at  the  Mass  Spectrometer.  Connect 

Gases1  & 2

 with 1/8” 

tubing to the ¼” Tee Piece supplied with the 2-off ¼” BSPT x 1/8” Swagelok fittings. Connect the inlet 

branch of the tee to the 

Source Gas

 connection at the rear of the generator using ¼” tubing as above.  

 

NB: - to minimize pressure losses position the Tee as close to the Mass Spectrometer as possible. 

 

 

                                      

 

 

 

 

 

 

Summary of Contents for NM20ZL

Page 1: ...ain Crescent Inchinnan Business Park Inchinnan Renfrew PA4 9RE Scotland UK Tel 44 0 141 812 8100 Fax 44 0 141 812 8200 info peakscientific com www peakscientific com USA 19 Sterling Road Suite 8 Bille...

Page 2: ...gen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 2 Document Change History Rev No Changed Initials Date 1 Document Created SK 20 09 07 2 Included new design of c...

Page 3: ...Allow the company to inspect the Goods and or any installation and any relevant packaging as and when reasonably required by the Company 5 Subject to the Company being notified of any defect as is ref...

Page 4: ...ZL User Guide Issue 2 4 Contents Page No Document History 2 Warranty Statement 3 Safety Notice To Users 3 Introduction 5 Unpacking Installation 5 Location 5 Air Connection 6 Principle of Operation 7 C...

Page 5: ...ectrometer it is supplying If this is not convenient then the generator can be sited elsewhere however consideration should be made to the lengths of pipe runs as pressure drops can result from extend...

Page 6: ...the unit Fittings are provided to connect these ports to the gas connections of the Mass Spectrometer A kit of fittings for connecting to the Mass Spectrometer is provided with this User Guide LC MS w...

Page 7: ...oves bulk moisture and harmful particles The air is now divided into 2 streams The first stream is passed through the Nitrogen Membrane to generate a flow of Nitrogen for the Curtain Gas output The Ni...

Page 8: ...or is such that it can deliver up to 26 Litres min Nitrogen Air at 110 psig with an additional 12 Litres min of Nitrogen at 80 psig Should the demand for gases be less than the rated output flow or in...

Page 9: ...ver as with all scientific and technical equipment it should be regularly inspected and serviced as below BA Inlet Filters Every 12 months RAC Filter Elements Every 12 months 1 BA inlet Filter Element...

Page 10: ...ter The Element is as shown and should be changed at 12 month intervals The Air supply to Generator MUST be turned OFF and the Generator MUST be de pressurised prior to attempting to remove ANY filter...

Page 11: ...psig 10 0 Barg Minimum Air Inlet Flow Rate 90 l min ATP Outlet Gas Maximum Pressure Drop Outlet Inlet P 8 psig Maximum Gas Outlet Pressure 110 psig Maximum Source Gas Outlet Flow Nitrogen Air 26 Litre...

Page 12: ...PEAK Scientific Instruments Ltd High Purity Nitrogen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 12 9 Pneumatic Diagram...

Page 13: ...Scientific Instruments Ltd High Purity Nitrogen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 13 Maintenance Log Model NM20ZL Serial number Work Done Remarks Date...

Page 14: ...PEAK Scientific Instruments Ltd High Purity Nitrogen Generator Instructions for use Manual 20th September 2007 NM20ZL User Guide Issue 2 14 Notes...

Reviews: