ABN2ZA
User Manual
Page 10 Peak Scientific – Rev 9A – 16/09/15
Installation
Generator Environment
The Generator is designed for indoor use only. It should be
installed adjacent to the Mass Spectrometer it is supplying. If
this is not convenient then the unit can be sited elsewhere,
however, consideration should be made of the lengths of pipe
runs as pressure drops can result from extended runs of pipe.
Please see the
Tubing lengths
section for further details.
Performance of the Generator (like all sophisticated
equipment) is affected by ambient conditions. Note should
also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high
relative humidity. Operation of the unit within such a pocket
could adversely affect its performance. Consideration should
also be given to the air flow around the unit. It is
recommended that an air gap of 75mm (3”) should be
maintained down both sides, at the rear and across the top of
the unit. Please refer to the drawing below for the general
dimensions of the unit.
Maximum Operating Ambient Conditions: 35°C (dry bulb) 80%RH (Max)
General Dimensions
Figure 1: General Dimensions
The Generator must always be placed on a level surface.
Failure to do so will affect the performance of the Generator.
LH
RH
600mm
750mm
712mm