Page 34
4. The deioniser column is held in place by two Velcro straps, these must be unfastened.
5. Next disconnect the deioniser column by pressing the silver tabs on the fittings at the
top and bottom of the column.
6. Remove the used column by pulling it towards you.
7. Fit a new deioniser column (05-7062) into the generator and reconnect the tubing
8. Refasten the Velcro straps and re-attach the front fascia, making sure to connect the
HMI screen to the unit again.
9. Finally, the HMI screen will display D.I. COLUMN CHANGED? The CONFIRM button
should now be selected to resume normal operation.
Summary of Contents for Precision Hydrogen 1200cc
Page 38: ...Page 38...
Page 39: ...Page 39...