1.
Electrical requirements for 230V generators is 110VAC 50/60Hz OR 230VAC
50/60Hz.
2.
The Teflon tube supplied in the fittings kit is to allow the generator to be installed
adjacent to the instrument it is supplying. If the generator is to be installed further
away from the instrument then attention needs to be paid to pressure drops that could
occur. The maximum distance away from the instrument the generator can be placed
still using 6mm (4mm internal diameter) or ¼” (3/16” internal diameter) tubing is 10m.
For further information relating to tubing lengths please refer to the user manual.
Electrical Requirements
Generator
Voltage
Frequency
Plug Type
230
110
50
60
MS Table
UK, EU, US 230V,
US 110V
Please familiarise yourself with the full
User Manual
available at
www.peakscientific.
com/downloads
.
This includes all other technical specifications, operating
instructions, service requirements, contact details and trouble shooting.