background image

Genius3045 User Manual Rev 2 RSID 127 PN UM-0018 17/11/21

Genius 3045

User Manual

Summary of Contents for Genius 3045

Page 1: ...Genius3045 User Manual Rev 2 RSID 127 PN UM 0018 17 11 21 Genius 3045 User Manual...

Page 2: ...12 Fittings Kit Contents 13 Installation 14 Generator Environment 14 Generator Overview 15 General Dimensions 15 Electrical Connection 17 Start Up Sequence 18 Normal Operation 19 On Demand Gas 19 Gene...

Page 3: ...document where you can skip to the relevant information Users can refer to the contents page to find the relevant information Please review each of the following sections carefully Thank you for sele...

Page 4: ...Intelligent High duty service indicator Mobile Flexibility to position the generator where required Simple installation Generator designed as a plug and play system Economical More cost effective than...

Page 5: ...10 Class 1 2 1 at all times d your air compressor dryer filtration and oil removal systems must be deemed suitable for use by Peak or a Peak Partner and must be changed and serviced regularly in line...

Page 6: ...y respect or to create any additional obligations or liabilities for any party involved 11 This warranty statement supersedes any and all prior warranty agreements between the parties and constitutes...

Page 7: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Page 8: ...e requirements Low Voltage Directive 2014 35 EU EN 61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Electromagnetic Compatibility Directive 2014 30...

Page 9: ...quipment Safety Regulations 2016 SI 2016 1101 as amended BS61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use The Electromagnetic Compatibility Regula...

Page 10: ...Guidance Notes PDF produced by the Department for Business Innovation and skills for the UK and here for Europe All PEAK products that are subject to the WEEE directive are compliant with the WEEE mar...

Page 11: ...low Rate 32 L min 1 13 cfm Maximum Air Outlet Flow Rate 50 L min 1 76 cfm Start up time 1 Hour Particles NONE Phthalates NONE Suspended Liquids NONE Gas Outlets 2 x BSPP Drain Outlets 1 x BSPP Pressur...

Page 12: ...e Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvr...

Page 13: ...Kit are as follows 1 6mm Teflon Tubing x 6m 2 Teflon Tubing x 6m 3 6mm PE Tubing x 3m 4 Compression Fitting x 3 5 6mm Push Fit Fitting x 3 6 Flow Control Silencer x 2 7 UK Mains Power Cable x 1 8 EU M...

Page 14: ...s affected by ambient conditions Note should also be taken to the proximity of Air Conditioning outlets These can sometimes give rise to pockets of air with high relative humidity Operation of the uni...

Page 15: ...15 Generator Overview General Dimensions The generator must always be placed on a flat level surface Failure to do so will affect the performance of the generator WARNING 600 mm 23 6 850 mm 33 5 1322...

Page 16: ...Page 16...

Page 17: ...s classified as SAFETY CLASS 1 THIS UNIT MUST BE EARTHED Before connecting the unit to the mains supply please check the information on the serial plate The mains supply must be of the stated AC volta...

Page 18: ...put pressure gauges on the front panel This will climb to the factory set pressure as noted in the specifications Once this pressure is reached the compressors will continue to run until the internal...

Page 19: ...arts again the system will detect the demand for gas and will automatically start again to suit the demand Generator Cycling The Generator is designed for the internal compressors to cycle This cyclin...

Page 20: ...e application can lead to a reduction in efficiency Tubing Lengths The diameter of the tubing which will be connected to the gas outlet is important and is determined by the length of tubing required...

Page 21: ...and offers a costs saving over buying the components separately Purchase Interval Component Part No Qty 12 months Genius 3045 Annual Service Kit 08 4787 1 Service Indication The generator will notify...

Page 22: ...noted that the effect of these extremes can force the compressors to run continually The compressors are fully capable of running continually with no detrimental effect however users may find the gene...

Page 23: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Page 24: ...e cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cle...

Page 25: ...ressure gauges are showing normal pressure Contact your service provider The mass spec is reporting low pressure Contact your service provider Yellow SERVICE LED on front panel is on constantly A comp...

Page 26: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Page 27: ...Page 27...

Page 28: ...ndustry leading rapid response service to our customers With Peak Protected your laboratory s productivity becomes our top priority To discuss Peak Protected generator cover and payment options speak...

Reviews: