background image

Notes:
Symbol

Meaning

Class II equipment

Pch Life TENS & PMS is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer 

or the user of the Pch Life TENS & PMS can help prevent electromagnetic interference by maintaining a minimum distance between portable 

and mobile RF communications equipment (transmitters) and the Pch Life TENS & PMS as recommended below, according to the maximum 

output power of the communications equipment.

Rated maximum output
power of transmitter
(W)

Separation distance according to frequency of transmitter
(m)

150 KHz to 80 MHz

P

d

167

.

1

80 MHz to 800 MHz

P

d

167

.

1

800 MHz to 2.5 GHz

P

d

333

.

2

0.01

0.117

0.117

0.233

0.1

0.369

0.369

0.738

1

1.167

1.167

2.333

10

3.689

3.689

7.379

100

11.667

11.667

23.333

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can 

be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating 

of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1    At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

Issue date of the manual: 12-25-2014

Production date: 12-30-2014

Batch: xxx-xxx

Type BF applied part

Manufacturer

30

31

4) Recommended separation distances between portable and mobile RF 

communications equipment and the Pch Life TENS & PMS

17. Date

III. Labels on the device

Operating instructions

Weight:39g

lnput:DC 3.7V

Size:86x43.2x10.6MM

Pch Life

Pch Life

Model:PCH Electrotherapy 

Stimulator System

Summary of Contents for PCHLIFE

Page 1: ...Power supply DC 3 7V Output voltage 42V 500 Output current 84mA 500 Consumed current 40mA Pulse width 100 S Frequency 1 110Hz Smallest are of electrode pad 4cm2 Timer 10 60 minutes...

Page 2: ...Troubleshooting 24 14 Disposal of the Unit 25 15 Warranty period 25 16 Electromagnetic Compatibility 25 17 Date 30 III Labels on the device 31 I Foreword Before operation please read this user s manu...

Page 3: ...electrical conductive medium The irritation can usually be reduced by using an alternate conductive medium like gel or alternate electrode placement 7 Pch Life TENS PMS must be kept out of the reach o...

Page 4: ...cription of the Device 6 1 General Description of the Device Pch Life TENS PMS is a portable and DC 3 7V battery powered multifunction device offering both Transcutaneous Electrical Nerve Stimulation...

Page 5: ...Rechargeable lithium battery 6 4 Mode Description We suggested that you initially experiment using each of the 6 modes The mode that gives you the most desirable sensations and comfort is the most ap...

Page 6: ...requency changes among 12 5Hz to 55 5Hz and there are 64 periods on 220ms off 100ms then the device stops for 1 second The device repeats this cycle all the time Mode 3 Mode 4 1 Timing 10 60 minutes 2...

Page 7: ...ith flammable gases 16 Never use this product in concurrence with high frequency surgical equipment it may result in burns at the site of the stimulator electrodes and possible damage to the control u...

Page 8: ...pplying the electrodes 5 Apply the whole surface of the electrodes firmly to the skin Do not use electrodes that do not stick properly to the skin or only partially stick to the skin 6 In case of skin...

Page 9: ...nd Mode 4 for 10 20 minutes twice or 3 times a day Figure 7 Mode 1 for 20 minutes and Mode 3 for 20 minutes twice or 3 times a day Figure 8 Mode 4 for 10 20 minutes and Mode 6 for 10 20 minutes twice...

Page 10: ...it becomes uncomfortable Always stay below that point of discomfort 3 Recommended application duration and Mode selection The modes to be used for improving the muscle performance are Modes 1 4 5 and...

Page 11: ...last approximately 1 hour When charging is finished the LCD will show a full battery cell 4 The battery should be charged for 10 hours or so before first use Notes Only charge the unit when battery i...

Page 12: ...store the electrodes in the reusable bag in a shady place Stockroom temperature 5 C 27 C 41 80 F and humidity of 30 80 No need to sterilize 3 Never apply the pads to any other surface other than your...

Page 13: ...of pad is not sticky Are you using pad when perspiring Use when not perspiring in a cool room Were the pads stored under high temperature high humidity or direct sunshine Replace the pad 25 24 14 Dis...

Page 14: ...meters do not deviate the essential performance requirement The device displays abnormally in LCD 2 As to patients The sensitivity of stimulation may be weaker or stronger but it does not produce safe...

Page 15: ...s equipment should be used no closer to any part of the Pch Life TENS PMS including cables than the recommended separation distance calculated from the equation applicable to the frequency of the tran...

Page 16: ...23 333 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the trans...