If the current measurement value compared against the reference vibration exceeds the desired
tolerance, the device can perform a calibration which can be started by pressing the
OK
button
on the calibration screen and confirming the subsequent dialog.
The calibration is performed autonomously by the device and should take less than 5 seconds.
After the calibration was finished a pop-up is displayed on the screen with the hint that the
calibration was successful as confirmation. Afterwards a short delay the device returns to the
main measurement screen.
5.3
Settings
5.3.1
Units
The measurement unit can be switched between the metric system and imperial units.
5.3.2
Date & time
This menu is used to change the date and time which is displayed in the status bar at the top of
the screen. The date format can also be changed.
5.3.3
Display brightness
The screen backlight intensity can be adjusted from 10% to 100%.
5.3.4
Language
This menu is used to switch between different menu languages. Available languages include
English, German, French, Spanish, Italian, Dutch, Portuguese, Turkish, Polish, Russian, Chinese
and Japanese.
5.3.5
Auto power off
This option enables the auto power off function. Available time periods include 1 minute, 5 minutes
and 15 minutes. After the set time period has elapsed, the device will automatically turn off and
any button press will reset the timer. It is also possible to disable the auto power off function.
5.3.6
Device reset
This menu is used to reset the device to factory settings. Device settings are separate from the
calibration and each can be reset individually by selecting the corresponding menu item.
A reset of the device settings will load default measurement parameters and reset any remaining
menu option to the factory default values. A calibration which may have been performed
beforehand is retained.
Fig. 11 Confirmation dialog