background image

Manual 

 
 

 

7  SETUP 

To enter the settings, press and hold the On/Off button for 2 seconds in normal measuring mode. 
You can choose between the following settings: 
 
P1.0: Unit selection 
P2.0: AREA size 
P3.0: RH offset 
 
To switch between the settings, use the Up and Down buttons. 
To exit the setup, press and hold the On/Off button again. 

7.1.1  P1.0: Unit selection 

When  P1.0  is  selected,  press  the  Hold  button  to  enter  the  unit  selection.  By  pressing  the  Up  or  Down 
button, you can now  switch between metric and imperial units. The units include air velocity (m/s, fpm), 
temp. (°C, F), volume flow (cmm, cfm) and area size (cm², inch2). After finishing the selection, press the 
Hold button to return to setup selection. 

7.1.2  P2.0: AREA Size 

In P2.0 you can set the dimensions of the cross-sectional area for volume flow measurements. Press the 
Hold  button  to  enter  the  area  setting. The  lower  display  now 

shows “99999”  with the first digit flashing. 

Press  the  Up  button  to  change  the  selected  digit  and  press  the  Down  button  to  move  to  the  next  digit. 
Once you set up the area properly, press the Hold button to confirm the input and to  return to the setup 
selection. 

7.1.3  P3.0: RH offset 

In  P3.0 you  can  set  a  humidity  offset. To  do  so,  press  the  Hold  button  when  P3.0  is  selected.  Now  the 
upper  display  shows  the  flashing  offset value.  Press  the  Up  button  to increase  the  offset value  and the 
Down button to decrease it. Once you are finished, press the Hold button again to confirm your input and 
to return to the setup selection. 
 

8  Calibration 

8.1  RH Calibration 

The humidity calibration of this meter requires specific fixture and cannot be done by end users. Please 
contact the dealer for calibration service. 
 

 

Summary of Contents for PCE-VA 20

Page 1: ...ruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce i...

Page 2: ...unctions 7 6 1 Data Hold 7 6 2 MIN MAX 7 6 3 Averaging 8 6 3 1 Multi point average 8 6 3 2 Time average 8 6 4 Auto power off function 8 7 SETUP 9 7 1 1 P1 0 Unit selection 9 7 1 2 P2 0 AREA Size 9 7 1...

Page 3: ...be used by qualified personnel and repaired by PCE Instruments personnel There is no warranty of damages or injuries caused by non observance of the manual Don t use the device with wet hands Do not u...

Page 4: ...measured value 0 3 m s when above 20 m s Temperature 0 6 C Relative humidity 3 at 25 C else 5 Technical Data Fan diameter 10 cm Fan anemometer dimensions 269 x 160 x 51 mm Airflow cone dimensions Rou...

Page 5: ...ure dew point temperature 4 1 Display Display indication Meaning HOLD Freeze the current display AVG Averaging MAX MIN Highest and lowest values Low battery warning m s fpm Flow velocity display units...

Page 6: ...MIN and MAX values In average mode press to record data or start timed recording In setup mode press to enter data settings Switches to upper display Long press to go into average mode In average mode...

Page 7: ...ttach one of the optional cones to the meter The anemometer detects the cone automatically and shows a cone indication on the display After that the device will calculate the volume flow Note Make sur...

Page 8: ...e current reading Press the Hold button to start recording The timer on the upper display starts counting Press the Up button to stop and calculate the timed average You can see the duration on the up...

Page 9: ...nsions of the cross sectional area for volume flow measurements Press the Hold button to enter the area setting The lower display now shows 99999 with the first digit flashing Press the Up button to c...

Page 10: ...fan anemometer for repair E03 Temperature is above measuring range Acclimatize the fan anemometer at room temperature for 30 minutes If E02 is still displayed return the fan anemometer for repair E31...

Page 11: ...s above range Take measurements within range If E03 is still displayed return the fan anemometer for repair 9 2 5 Volume Flow Error code Error Remedial action E03 Value is above display range Check en...

Page 12: ...contact PCE Instruments 10 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire United Kingdom SO31 4RF By phone 02380 987 035 10 2...

Reviews: