Error codes and countermeasures
Error codes are indicated by a single letter in the lower right corner of the display. However, these
only occur in the menus M00, M01, M02, M03, M90 and M08. The following chart shows the error
codes and countermeasures.
Error
code
Message in window
M08
Reason
Countermeasure
R
System Normal
No error
- - -
I
Detect No Signal
(1) No signal
(2) Sensors installed
improperly
(3) Too much fouling
(4) Liner too thick
(5) Sensor cable not
Properly
connected
(1) Change
measuring
location
(2) Clean measuring
location
(3) Check the cables
J
Hardware Error
Hardwareproblem
Contact PCE
Instruments
H
PoorSig Detected
(1) Schlechtes Signal
(2) Sensoren falsch
angebracht
(3) Zu viel Bewuchs,
zu starke
Verschmutzung
(4) Auskleidung vom
Rohr zu dick
(5) Sensorkabel nicht
korrekt
angeschlossen
(1) Change
measuring
location
(2) Clean measuring
location
(3) Check the cables
(4) Check the
contact gel
Q
Frequ OutputOver
Die
Ausgangsfrequenz
liegt außerhalb des
erlaubten Bereiches
Check the values in
the windows M67,
M68 and M69. Enter
higher values in
window M69.
F
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
(1) Vorübergehende
Probleme mit dem
RAM oder RTC
(2) Permanente
Probleme mit der
Hardware
(1) Restart the device
(2) Contact PCE
Instruments
1
2
3
Adjusting Gain
Das Gerät stellt
gerade die
Signalverstärkung
(Gain) neu ein; die
Zahl gibt den
aktuellen
Arbeitsfortschritt an
- - -
K
Empty pipe
(1) Keine Flüssigkeit
in der Rohrleitung
(2) Einstellungsfehler
im Menü M29
(1) Choose a pipe
that contains
liquid
(2) Enter
“0“ in
window M29.