background image

 

 

©  PCE Instruments 

11 

 

10  Alarm 

Inter:

 This option triggers an audible signal when the measured value is between the 

lower and the upper alarm limit. 

„+“ and „-“ will flash in the display. 

 

Outer: 

This option triggers an audible signal when:

 

 

A: the measured value is below the lower limit 

(„-“ will flash in the display) 

 
B: the measured value is above the upper limit 

(„+“ will flash in the display) 

 

Off: 

This option deactivates the alarm.

 

11  Calibration by using weights 

You can select either 

„cal=2“ (2-point calibration) or „cal=3“ (3-point calibration). 

 
Go to the menu F-3 and either select 

„cal=2“ or „cal=3“ by means of the arrow keys. Confirm your 

selection with OK. Remove all attachments from the device that affect the measuring cell. Confirm 
the zero calibration by pressing the OK key. To calibrate the second point, you can use the pre-
set weight or select the desired weight by using the arrow keys. Confirm with the OK key. Attach 
the  chosen  weight  to  the  device  and  confirm  with  OK.  The  indication  in  the  display  will  start 
flashin

g. When the calibration is finished, the display will “CAL”. 

 
The procedure for the 3-point calibration is the same as for the 2-point calibration, just with one 
more calibration point. 

12  Rechargeable battery 

This product is configured with a 1600 mAh 6 V Ni-Hi rechargeable battery. If it is fully charged, 
the product can be running uninterruptedly for 10 hours. When shut down, the battery will 
discharge in 3 months. Make sure the battery level is always sufficient. In case of power 
shortage, use the delivered DC 12 V/1000 mA power adaptor to charge the product. It will be 
fully charged within 8 ~ 10 hours. Remove the power adapter immediately after the product 
prompts full charge. Otherwise, longtime charging may cause the battery to overheat. Only 
charge the battery when it is completely flat. Frequent charging shortens the battery life. 

13  Measurement 

13.1 

Choosing the sample 

Following this procedure before making a measurement is very important to ensure valid readings 
which  are  helpful  when it comes to  determining the  best  harvest time.  Make sure the  samples 
represent a typical part of the total harvest (minimum number of samples, sizes, quality, etc.). It 
is recommended to create your own sampling protocol. The samples should come from various 
areas of the harvest. It is important that the selected fruit has equal average sizes as the firmness 
also depends on the size (the firmness increases with the size). I is also crucial that the chosen 
samples are completely flawless and undamaged. The measuring interval should be about every 
3 … 5 days and approx. 10 pieces of fruit should be measured. 
 

 

Summary of Contents for PCE-PTR 200N

Page 1: ...PCE PTR 200N Dynamometer User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 2...

Page 2: ...ption 4 7 1 Keys and adaptor 4 7 2 Dimensions 5 7 3 Circuit diagram of 3 5 mm jack 5 7 4 Display 6 7 5 Keys 7 8 Operation 8 9 Options 10 10 Alarm 11 11 Calibration by using weights 11 12 Rechargeable...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...applied is smaller than 105 the rated load after the meter is restarted Make sure the force gauge is not overloaded Overload may damage the sensor Do not use sharp tools to press buttons Keep the met...

Page 5: ...for a test stand This reduces possible user error to a minimum and makes series of measurements possible 5 Characteristics Display LCD 180 rotatable backlit Interface USB Alarm Fracture alarm upper l...

Page 6: ...PCE Instruments 4 7 Device description 7 1 Keys and adaptor 1 Measuring adaptor 2 Clamping nut 3 LCD 4 LED 5 Control panel 1 Switched output 2 USB interface 3 Power connection...

Page 7: ...truments 5 7 2 Dimensions All dimensions in mm 7 3 Circuit diagram of 3 5 mm jack Here you can find the circuit diagram for connecting a 3 5 mm jack in order to establish a connection to the switched...

Page 8: ...ighest value measured MIN lowest measured value 3 Displays current the peak value or current measurement value in real time depending on display orientation 4 Battery level indication 5 PC connection...

Page 9: ...tion Zeroing Zeroing Memory is cleared x Reading and unit are sent x Data are sent Up Saved readings are qeried x Switch between AVER MAX and MIN Down Menu is opened Invalid Back to measurement mode S...

Page 10: ...ndividually Change unit and leave a mode In test mode press to change the unit In the menu you can use to leave a selected option or menu In storage mode you can use to leave this mode and return to t...

Page 11: ...ing saved values AVER MAX MIN In test mode you can open storage mode with the key In storage mode you can press to switch between AVER MAX and MIN USB interface Please observe the instructions in the...

Page 12: ...arm deactivated outer outside alarm limit F 7 lo Lower alarm limit 0 F 8 hi Upper alarm limit Maximum F 9 cut 10 90 Overload alarm 50 F 10 peak key 3 60 sec Display time of peak value Remove by pressi...

Page 13: ...attery This product is configured with a 1600 mAh 6 V Ni Hi rechargeable battery If it is fully charged the product can be running uninterruptedly for 10 hours When shut down the battery will discharg...

Page 14: ...ecisive for the firmness of fruit i e the higher the temperature the softer the fruit will be 13 3 Important information about the measurement After preparing the sample and selecting the right tip yo...

Page 15: ...ation tip 4 Peeled zone Try to make all measurements under equal conditions to ensure reproducible values for comparative and statistical purposes To achieve high accuracy use a test stand which enabl...

Page 16: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 17: ...583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info pce instruments co uk info pce italia it www pce instruments com deutsch www pce instrumen...

Reviews: