-  to increase the battery life you should often  use it 

   

-  remember about correct polarity plug of power supply fig. 2       

 

 

Fig. 2 Correct polarity the plug of power supply 

 

4. Additional features include: remembering of the last measurement result , counting, additional 
filtration 
        Remembering  of the  of the last weighing

 

result 

 

-

 

press the button "F" (F1) (the inclusion of this feature is indicated by a flashing sign "+" or 
"-" in front of the weight value); 

-

 

after action of  loading the pan, transient time is indicated by  flashing  

-

 

stabilize indications cause blocks of the result and a stable reading of it, even after remov-
ing the load;  

-

 

if you want to cancel the result of the weighing press the "TARA"  button. 

-

 

Counting the elements of the same mass.

-  

The scale can determine the mass per unit calculated on the basis of sample components with 
different frequencies: 1 pc., 2p., 5p., 10p., 20p., 50p., 100p., 200p., 500p. The system has the 
limitation  of  software  to  prevent  counting  calibration  sample  with      mass  <1%  Max.    For 
counting heavier and  less numerous  elements  you should calibrate the procedure of counting 
using  a  smaller  sample  (eg,  10  units).  During  counting  lighter  and  more  numerous  elements 
you can calibrate the procedure using  more numerous of samples  (eg, 100).- The counting er-
rors is affected by  many factors. The most important include: the appropriate selection of the 
range and accuracy of the scale, determine the calibration  of sample size, the differences be-
tween the mass of the individual component 

 

-

 

to start counting function repeatedly press "F" (F1) until the display indicates sign "Count";  

-

 

 then press the "TARA" buitton for approval of choice-window; it will show for example 

an indication "L10", where number is the sample size, which can be changed by pressing "F" 
(F1);  

-

 

after    pressing  the  button  "F"  (F1)  you  choose  adequately  large  sample  (see  description 

above); wait about 3 sec for the automatic change display indication  You should see a flash-
ing display "I_ ___ I". No change the indication means that the reading is unstable; 

-

 

put on the  pan deductible items; 

-

 

continuous indication "1___ _ I" means that the minimum sample weight was achieved (for 

very small mass calibration of the counting is not possible);  

-

 

press the "TARA" button to accept the calibration mass;  

-

 

appearance of letters indication “ps” means  the completion of the calibration procedure of 

counting.  Number of pieces  is shown on the display of scale. In counting mode tare function 
works  in  the  same  way  as  in  the  weighing  mode.  Analog  line  does  not  work  in  counting 
mode.  

-

 

subsequent press "F" (F1) – start the menu to  select the level of filtration (

additional fil-

tration

 can weigh moving objects) 

 

 