background image

 

 

 

©  PCE Instruments 

 

Operation 

Remove the black protective cap from the device and optionally unscrew the electrode cap if you 
want to use a different electrode. To change the electrode, you can simply remove it and attach 
another one. 
 

Note

: Do not touch the sensors of the electrodes!  

5.1 

pH calibration 

Since the electrode is subject to a time drift, it should be calibrated regularly. The interval depends 
on the intended use and the required accuracy. 
 

 

Make sure that the pH electrode is mounted and the pH icon is shown on the display.  

 

Clean the electrode with distilled water.  

 

Immerse the electrode in the calibration solution pH 7, stir carefully and wait until the 
value has stabilized. 

 

Press and hold the CAL key until "Cal" and the calibration value are displayed. 

 

Calibration  is  performed  automatically  and  is  finished  when  "End"  is  displayed.  The 
instrument returns to measuring mode. 

 

Clean the electrode with distilled water. 

 

Now repeat the calibration with the pH 4 calibration solution.  
 

Notes: 

 

 

Always carry out a two-point calibration. Start at pH 7, followed by pH 4 (or 10). 

 

When  not  using  the  pH  electrode,  always  keep  the  electrode  moist  with  the  storage 
liquid. 
 

5.2 

Calibrate conductivity 

1. 

Make  sure  that  the  conductivity  electrode  is  mounted  to  the  instrument.  The  display 
shows COND, TDS or Salt.  

2. 

Immerse the conductivity electrode in the 1413µS/cm calibration solution and wait until 
the measured value has stabilized. 

3. 

Press and hold the CAL button until "Cal" is displayed and the calibration value flashes. 
Calibration  is  performed  automatically  and  is  finished  when  "End"  is  displayed.  The 
instrument returns to measuring mode. 

Notes: 

 

 

A calibration with 12.88mS/cm can also be performed optionally. It is suitable for 
measurements with high conductivity. 

 

If the measured value is not 0µS/cm in air after calibration, calibrate the instrument in 
air. 

5.3 

pH measurement 

1. 

Clean the pH electrode with distilled water.  

2. 

Dip the electrode tip into the liquid to be examined and move the tip slightly back and 
forth.  

3. 

Wait briefly until the display shows a stable pH value. 

 

 

Summary of Contents for PCE-PH 30

Page 1: ...30 pH Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 14...

Page 2: ...1 3 Specifications 2 3 1 Delivery contents 3 4 Device 3 5 Operation 4 5 1 pH calibration 4 5 2 Calibrate conductivity 4 5 3 pH measurement 4 5 4 Conductivity measurement 5 5 5 Redox measurement 5 6 K...

Page 3: ...when your hands are wet You must not make any technical changes to the device The appliance should only be cleaned with a damp cloth Use only pH neutral cleaner no abrasives or solvents The device mus...

Page 4: ...e ATC 0 50 C 32 122 F Calibration 0 S 1413 S 12 88 mS Measuring function TDS Measurement range 0 1300 ppm 1 3 13 ppt Resolution 1 ppm 0 01 ppt Accuracy 2 of measurement range ATC 0 50 C 32 122 F Measu...

Page 5: ...calibration liquid 1 x pH 7 calibration liquid 1 x 1413 S cm calibration liquid 1 x storage solution 1 x wrist strap 4 x 1 5 V AAA battery 1 x user manual 4 Device Battery compartment Display pH elect...

Page 6: ...at the calibration with the pH 4 calibration solution Notes Always carry out a two point calibration Start at pH 7 followed by pH 4 or 10 When not using the pH electrode always keep the electrode mois...

Page 7: ...he liquid to be measured 3 Wait for a short time until a stable value is displayed Notes The display shows if the value exceeds the measurement range When not using the electrode always keep the elect...

Page 8: ...terms 9 Disposal For the disposal of batteries in the EU the 2006 66 EC directive of the European Parliament applies Due to the contained pollutants batteries must not be disposed of as household was...

Page 9: ...e stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co...

Reviews: