© PCE Instruments
4
5
Operation
Remove the black protective cap from the device and optionally unscrew the electrode cap if you
want to use a different electrode. To change the electrode, you can simply remove it and attach
another one.
Note
: Do not touch the sensors of the electrodes!
5.1
pH calibration
Since the electrode is subject to a time drift, it should be calibrated regularly. The interval depends
on the intended use and the required accuracy.
•
Make sure that the pH electrode is mounted and the pH icon is shown on the display.
•
Clean the electrode with distilled water.
•
Immerse the electrode in the calibration solution pH 7, stir carefully and wait until the
value has stabilized.
•
Press and hold the CAL key until "Cal" and the calibration value are displayed.
•
Calibration is performed automatically and is finished when "End" is displayed. The
instrument returns to measuring mode.
•
Clean the electrode with distilled water.
•
Now repeat the calibration with the pH 4 calibration solution.
Notes:
•
Always carry out a two-point calibration. Start at pH 7, followed by pH 4 (or 10).
•
When not using the pH electrode, always keep the electrode moist with the storage
liquid.
5.2
Calibrate conductivity
1.
Make sure that the conductivity electrode is mounted to the instrument. The display
shows COND, TDS or Salt.
2.
Immerse the conductivity electrode in the 1413µS/cm calibration solution and wait until
the measured value has stabilized.
3.
Press and hold the CAL button until "Cal" is displayed and the calibration value flashes.
Calibration is performed automatically and is finished when "End" is displayed. The
instrument returns to measuring mode.
Notes:
•
A calibration with 12.88mS/cm can also be performed optionally. It is suitable for
measurements with high conductivity.
•
If the measured value is not 0µS/cm in air after calibration, calibrate the instrument in
air.
5.3
pH measurement
1.
Clean the pH electrode with distilled water.
2.
Dip the electrode tip into the liquid to be examined and move the tip slightly back and
forth.
3.
Wait briefly until the display shows a stable pH value.