Specifications
Current simulation (active)
(max. burden 1 kOhm)
Area
Resolution
Accuracy
0 ... 22-mA
0.001-mA
± 0.05% of sim. value + 4 µA
Current simulation (passive)
(max. burden 1 kOhm / max. 5 ... 25V DC)
Area
Resolution
Accuracy
0 ... 22-mA
0.001-mA
± 0.05% of sim. value + 4 µA
Voltage transmitter f. current
loop
Area
Resolution
Accuracy
24V DC / 25-mA
± 10%
Voltage measurement
Area
Resolution
Accuracy
-0.2 ... 28V
1 mV
± 0.02% of meas. + 2 mV
Current measurement
Area
Resolution
Accuracy
-1 ... 22-mA
0.001-mA
± 0.02% of meas. + 4 µA
Current measurement
(Loop current with voltage transmitter function)
Area
Resolution
Accuracy
0 ... 22-mA
0.001-mA
± 0.02% of meas. + 4 µA
More specifications
Power supply
2 x 1.5V AA batteries
Max. Tension
30V
Operating conditions
0 ... 50°C / 32 ... 122°F, < 80% RH
Storage conditions
-10 ... 60°C / 14 ... 140°F, < 95% RH
Dimensions
180 x 90 x 47 mm / 7.1 x 3.5 x 1.9 in
Weight
Approx. 500 g / 1.1 lbs