background image

Remark :

The adjusted unit will be flashed.

2)After set all the time value ( Year, Month, Date, Hour,

Minute, Second ), press the  " SET button " ( 3-5, Fig.
1 ) once will save the time value,  then the screen will
jump to  Sampling time " setting screen ( Chapter 7-3 ). 

Remark :

After the time value is setting, the internal clock will

run precisely even Power off if the battery is under

normal condition ( No low battery power ).

7-3 Set sampling time 

When the Display show " SP-t " 

1)Use the "   Button " ( 3-3, Fig. 1 ) or  "   Button " 

( 3-4, Fig. 1 ) to adjust the sampling value :

5 seconds, 10 seconds, 30 seconds, 60 second,

120 seconds, 300 seconds, 600 seconds Auto.

After the desired value is set,  press the   " Enter Button "
 ( 3-2, Fig. 1 ) to save the adjusting value with default.

Remark :

The " Auto " sampling time means when the measuring 

value is changed  ( > ±  1 %RH or  > ±  1   ) will save

the data to the memory circuit one time.

7-4 Set beeper sound ON/OFF

When the Display show " bEEP " 

1)Use the "   Button " ( 3-3, Fig. 1 ) or  "   Button "  

( 3-4, Fig. 1 ) to select the data  to " yES " or " no ".

Summary of Contents for PCE-HT 110

Page 1: ...Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southam...

Page 2: ...nd 4 3 9 Battery cover Battery compartment 4 3 10 Screw of the battery cover 4 3 11 Reset button 4 3 12 RS 232 output terminal 4 3 13 SD card socket 4 3 14 DC 9V power adapter input socket 4 3 15 Hang...

Page 3: ...the precision capacitance type humidity sensor professional and high accuracy SD card capacity 1 GB to 16 GB Sampling adjustment 5 10 30 60 120 300 seconds and auto function Large LCD display easy rea...

Page 4: ...onnect the optional RS232 cable UPCB 02 will get the RS232 plug Connect the optional USB cable USB 01 will get the USB plug Operating 0 to 50 Temperature Operating Less than 85 R H Humidity Power Supp...

Page 5: ...n 0 1 R H Accuracy 70 RH 3 reading 1 RH 70 RH 3 RH Temperature Range 0 to 50 32 to 122 Resolution 0 1 degree Accuracy 0 8 1 5 Above specification tests under the environment RF Field Strength less tha...

Page 6: ...ry UM4 AAA Alkaline Heavy duty type x 6 PCs and reinstate the cover Make sure the battery cover is secured after changing the battery 2 The Display 3 1 Fig 1 will show both the humidity and the temper...

Page 7: ...mal character at first details of setting the Decimal point refer to Chapter 7 5 page 13 e 3 Information of LCD display If the Display show It means that the SD card exist the problem CH or the SD car...

Page 8: ...h the Datalogger During execute the Datalogger function Display show the Datalogger indicator press the Logger button 3 2 Fig 1 2 seconds continuously until the Display indicator DATALOGGER is disappe...

Page 9: ...After exist the Datalogger then execute again the data will save to the HTC01001 XLS until Data column reach to 30 000 columns then will generate a new file for example HTC01002 XLS 3 Under the folde...

Page 10: ...e SD card into the SD card adapter then connect the SD card adapter into the computer 3 Power ON the computer and run the EXCEL software Down load the saving data file for example the file name HTA010...

Page 11: ...n release the SET button Following press the SET button 3 5 Fig 1 once a while in sequence to select the seven main function the display will show Sd F SD memory card Format dAtE Set clock time Year M...

Page 12: ...Fig 1 once again the Display will show text yES Enter to confirm again if make sure to do the SD memory card format then press Enter Button once will format the SD memory clear all the existing data...

Page 13: ...power 7 3 Set sampling time When the Display show SP t 1 Use the Button 3 3 Fig 1 or Button 3 4 Fig 1 to adjust the sampling value 5 seconds 10 seconds 30 seconds 60 second 120 seconds 300 seconds 60...

Page 14: ...a structure of SD card is used the as the decimal with default for example 20 6 1000 53 But in certain countries Europe is used the as the decimal point for example 20 6 1000 53 Under such situation i...

Page 15: ...will save the setting function with default 7 7 Set RS232 data output ON OFF When the Display show rS232 1 Use the Button 3 3 Fig 1 or Button 3 4 Fig 1 to select the upper Display text to yES or no yE...

Page 16: ...2 Loose the Screw of the battery cover 3 10 Fig 1 and take away the Battery Cover 3 9 Fig 1 from the instrument and remove the battery 3 Replace with DC 1 5 V battery UM4 AAA Alkaline heavy duty x 6...

Page 17: ...lead with the following connection will be required to link the instrument with the PC serial port Meter PC 9W D Connector Center Pin Pin 4 3 5 mm jack plug Ground shield Pin 2 2 2 K resistor Pin 5 Th...

Page 18: ...ata 1 When send the lower display data 2 D14 4 D15 Start Word RS232 FORMAT 9600 N 8 1 Baud rate 9600 Parity No parity Data bit no 8 Data bits Stop bit 1 Stop bit 12 PATENT The meter SD card structure...