© PCE Instruments
4
No.
Description
1
LC display
2
Setup key
3
On/off key
4
Mode key
5
Micro USB port
5
Switch on/off
To switch on the meter, press and hold the on/off key for 2 seconds. Make sure you switch on the
meter in fresh air as the meter calibrates itself in the first 90 seconds. This procedure is always
performed when the meter is switched on.
Leave the meter in fresh air for another 5 ... 10 minutes so that it can adapt to the environment.
The temperature and humidity sensors also need 20 ... 30 minutes to adapt to the environment.
To switch off the meter, press the on/off key again for 2 seconds.
6
Backlight
To switch on the backlight, press and release the on/off key once. To switch off the backlight,
press the on/off key again. When the measured value is >0.09 mg/m³, the backlight turns red. As
soon as the measured value is below >0.09 mg/m³, the backlight turns white again. Even if the
backlight is disabled, the red lighting switches on when the value is exceeded.
7
Change temperature unit
To change the temperature unit, press and release the setup key once. The temperature unit can
be set to either °C or °F.
8
Highest measured value
From the moment the unit is switched on, the highest measured value is temporarily saved. To
have this measured value displayed, press the mode key. The display shows "MAX" and the
highest measured value is shown on the display. To exit this display, press the mode key again.
To reset this display without restarting the meter, press and hold the mode key for 3 seconds.