5  Instructions 

 

DP = dew point 
 
The dew point or dew point temperature is defined as the temperature at which condensing and 
evaporating water on an object (when there is humidity) are at equilibrium or, in other words, at which 
condensation starts. 
 
AT = ambient temperature 
 
The device measures the ambient temperature in your environment and indicates it on the display. 
 
After you switch on the device, ambient temperature, humidity and dew point temperature are measured 
and indicated on the display. Press the MODE key to switch between temperature, humidity and dew 
point temperature. 

 

Target at the spot you wish to measure. Remove the detector slowly to determine the temperature of the 
object to be measured. If the measured temperature is between the ambient temperature and the dew 
point temperature, the meter will show the simulated D value with the ambient temperature. 
 
The meter can detect mould if the measured temperature value is in the first third of the range between 
ambient temperature and dew point temperature. The green LED indicates that temperature and humidity 
are within the normal range. 
 
If the measured temperature value is in the second third of the range between ambient temperature and 
dew point temperature, the yellow LED indicates that mould may form at this spot. 
 
If the measured temperature value is above the second third of the range between ambient temperature 
and dew point temperature, the red LED indicates that mould has already formed at this spot. 

Summary of Contents for PCE-DPT1

Page 1: ...ricas com www pce instruments com us PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Southampton Hampshire United Kingdom SO31 4RF Tel 02380 98703 0 Fax 02380 98703 9 info indus...

Page 2: ...12 2014 Date of last change 30 12 2014 Table of contents 1 Introduction 3 2 Safety information 3 3 Specifications 4 4 Device description 4 5 Instructions 5 6 Maintenance 6 7 Disposal 6 8 Contact 7 8 1...

Page 3: ...nual The device may only be used in approved temperature and humidity ranges as stated in the specifications Extreme temperatures direct exposure to sunlight extreme atmospheric humidity or moisture e...

Page 4: ...ed at 0 95 Temperature accuracy 3 5 C 50 20 C 58 32 F 1 1 5 C 20 350 C 32 662 F Wavelength range 8 14 m Humidity range 0 100 RH Humidity accuracy 3 5 20 80 Dew point temperature 30 100 C 22 212 F D S...

Page 5: ...or slowly to determine the temperature of the object to be measured If the measured temperature is between the ambient temperature and the dew point temperature the meter will show the simulated D val...

Page 6: ...the battery To do so open the battery compartment cover and replace the battery by a new 9 V battery 7 Disposal For the disposal of batteries the 2006 66 EC directive of the European Parliament appli...

Page 7: ...1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire United Kingdom SO31 4RF By phone Support 02380 987 035 Sales 02380 987 030 8...

Reviews: