background image

User Manual  

Pressure Gauge PCE-DMM 70 / PCE-DMM 60 

1.1 Information on the operating manual 

This  operating  manual  contains  important  information  on  proper 
usage of the device. Read this operating manual carefully before 
installing and starting up the pressure measuring device. 

Adhere  to  the  safety  notes  and  operating  instructions  which  are 
given in the operating manual. Additionally applicable regulations 
regarding  occupational  safety,  accident  prevention  as  well  as 
national  installation  standards  and  engineering  rules  must  be 
complied with! 

This operating manual is part of the device and  must be kept at 
a,  for  the  personnel  always  accessible  location,  close  to  the 
installation position of the device. 

This  operating  manual  is  copyrighted.  The  contents  of  this 
operating  manual  reflect  the  version  available  at  the  time  of 
printing.  It  has  been  issued  to  our  best  knowledge  and  belief. 
However,  errors  may  have  occurred.  For  incorrect  statements 
and  their  consequences,  liability  cannot  be  assumed  by  PCE 
Instruments. 

– Technical modifications reserved – 

1.2 Symbols used 

DANGER! 

–  Dangerous  situation,  which  may  result  in 

serious or fatal injuries 

WARNING! 

–  Potentially  dangerous  situation,  which  may 

result in fatal or serious injuries 

CAUTION! 

–  Potentially  dangerous  situation,  which  may 

result in minor injuries 

!

CAUTION! 

–  Potentially  dangerous  situation,  which  may 

result in damage of objects 
NOTE 

–  Tips  and  information  for  the  user  to  ensure  good

conditions for the operation. 

1.3 Target group 

WARNING! To avoid hazards for the operator and damages 
of  the  device,  following  described  instructions  have  to  be 
worked out by qualified technical personnel. 

1.4 Limitation of liability 

No liability is assumed and warranty claims are excluded in case 
of non-observance of the operating manual, improper application, 
modification of or damage to the device. 

1.5 Intended use 

- The  battery  powered  digital  gauge  has  been  designed  for 

extremely  high  demands  in  the  sector  of  calibration  and  test 
technology. It can be easily and quickly installed in situ. 

- It  is  in  the  responsibility  of  the  user  to  verify  whether  the 

chosen device is suitable for the intended application. In case 
of  any  doubts,  contact  our  sales  department  to  eliminate  any 
indistinctness. PCE Instruments does not assume any liability 
for an incorrect selection and its consequences! 

WARNING! - Danger through improper usage!  

1.6 Package contents 

Please verify that all listed parts are undamaged included in the 
delivery and check for consistency specified in your order. 

The  batteries  are  already  used.  The  circuit  is  interrupted  by  an 
insulation  foil.  Take  this  before  first  introduction,  see  in  addition 
battery change!! 

2. Mechanical Installation

2.1 Mounting and safety instructions 

WARNING!

 

Mount the device (pressure transmitter module) 

always in the state without pressure and apart from the dis-
play! 

WARNING!

 

This  device  may  only  be  installed  by  qualified 

technical  personnel  who  has  read  and  understood  the 
operating manual! 

WARNUNG!

 

Do  not  use  the  display  to  tighten  or  solve  to 

the mechanical connection of the pressure transmitter mod-
ule! 

!

Handle  this  electronic  precision  measuring  device  carefully 
in packed as well as in unpacked condition! 

!

The  device  must  not  be  subject  to  any  changes  or 
modifications. 

!

The device may not be thrown!

!

To  avoid  damaging  the  diaphragm,  remove  packaging  and 
protective cap only directly before starting up the device. A 
delivered protective cap must be stored! 

!

Place  the  protective  cap  on  the  pressure  port  again 
immediately after disassembling. 

!

Handle the unprotected diaphragm very carefully - it is very 
sensitive and may be easily damaged. 

!

Do  not  use any  force  when  installing  the  device  to  prevent 
damage of the device and the plant! 

Take  note  that  no  inadmissibly  high  mechanical  stresses 
occur  at  the  pressure  port  as  a  result  of  the  installation, 
since this may cause a shifting of the characteristic curve or
to  the  demage.  This  is  especially  important  for  very  small 
pressure ranges as well as for devices with a pressure port 
made of plastic. 
In hydraulic systems, position the device in such a way that 
the pressure port points upward (venting). 
Provide a cooling line when using the device in steam lines. 

2.2 General installation steps 

- Carefully  remove  the  pressure  measuring  device  from  the

package and dispose of the package properly. 

- Then  go  ahead  as  detailed  in  the  specific  instructions 

below. 

2.3 Installation steps for DIN 3852 

DO NOT USE ANY ADDITIONAL SEALING MATERIALS, 
LIKE YARN, HEMP OR TEFLON TAPE! 

- Check  to  ensure  the  proper  groove  fitting  of the  o-ring  and 

additionally to ensure no damage to the o-ring. 

- Ensure that the sealing surface of the taking part is perfectly 

smooth and clean. (R

3.2) 

- Screw the device into the corresponding thread by hand.
- If you have a device with a knurled ring, the transmitter has 

to be screwed in by hand only. 

- Devices  with  a  spanner  flat  have  to  be  fully  tightened  with 

an open-end Devices (G1/4": approx. 5 Nm; G1/2": approx. 
10 Nm). 

-

The  indicated  tightening  torques  must  not  be 
exceeded!

 

2.4 Installation steps for EN 837 

- Use  a  suitable  seal,  corresponding  to  the  medium  and  the

pressure input (e. g. a cooper gasket). 

- Ensure that the sealing surface of the taking part is perfectly 

smooth and clean. (R

Z

 6.3) 

- Screw the device into the corresponding thread by hand.
- Tighten  it  with  a  wrench  (for  G1/4":  approx.  20  Nm;  for

G1/2": approx. 50 Nm). 

-

The  indicated  tightening  torques  must  not  be 
exceeded!

 

2.5 Installation steps for NPT 

- Use a suitable seal (e. g. a PTFE-strip). 
- Screw the device into the corresponding thread by hand.
- Tighten it  with  a  wrench  (for  1/4"  NPT:  approx.  30  Nm;  for

1/2" NPT: approx. 70 Nm). 

-

The  indicated  tightening  torques  must  not  be 
exceeded!

 

2.6 Installation steps for internal threads M20x1.5 and 9/16" 

UNF (for DM01-500 HD) 

- Screw the high pressure connection into the internal thread 

of the pressure port and tighten it properly with approx. 160 
Nm. 

-

DANGER!  The  high  pressure  tube  seals  metal-to-metal 
in  the  chamfer  of  the  pressure  port.  No  further  seal  is 
allowed  with  this  high  pressure  connection.  A  wrong 
installation can cause enormous danger! 

3. Connection display with pressure transmitter module

Connect display with 

pressure transmitter

 module as follows: 

- bring  together  carefully  the  display  with 

pressure  transmit-

ter

 module. 

- press  the  display  sturdy 

pressure  transmitter

  module  to  this 

engages.

4. 

Supply

 / 

changing the batteries

As  soon  as  in  the  display  the  announcement  of  "battery"  is 
shown, carry out battery change as follows: 
- unscrew three fixing screws with a suitable screwdriver.
- take the battery case cap and exchange the batteries 3 x

1.5 V AA 

(remove the insulation foil before first intro-

duction)

- lock the device after that properly

!

An incorrect usage may cause a leak out of batteries and
so a damage the 

device

!

Never combine batteries of different types or old with new
ones! 

!

Make sure that the batteries are connected correctly with
the corresponding contacts in the battery tray.

!

Never try to charge batteries, demount them, or short -
circuit them. 

!

use only batteries with UL certification

!

Keep the batteries away from heat and unshielded  flame.

 

5. Data logger

 

The battery powered digital gauge disposes of an integrated data 
logger. The  measuring  values  stored  away  in  the  device  can  be 
selected  above  the  communication  interface  by  means  of 
software.

 

5.1 PC - connection 

Connect 

device

 with a computer as follows: 

- unscrew the protective screwing of the communication

interface with a suitable slit screwdriver.

- connect the handle plug of the connecting cable (included

in delivery) with the interface socket of the 

device

. Con-

nect the side with the USB plug with a free USB connec-
tion on the computer. 

- install the COM driver and data logger software.
- after the use, disconnect the connection and lock the

protection screwing again properly.

6. Initial start-up 

 

remove the insulation foil before first introduction

WARNING!

 

Before  start-up,  the  user  has  to  check  for 

proper installation and for any visible defects. 

WARNING!

 

The  device  can  be  started  and  operated  by 

authorized  personnel  only,  who  have  read  and  understood 
the operating manual!! 

WARNING!

 

The device has to be used within the technical 

specifications, only (compare the data in the data sheet)! 

7. Placing out of service 

WARNING!

 

When dismantling the device, it must always be 

done in the depressurized and currentless condition! Check 
also if the medium has to be drained off before dismantling! 

WARNING!

 

Depending  on  the  medium,  it  may  cause 

danger  for  the  user.  Comply  therefore  with  adequate 
precautions for purification. 

8. Operation 

8.1 Operating and display elements 

  

The  display  of  the  measuring  value  as  well  as  configuring  the 
single  parameters  occurs  menu-steered  about  a  LC  display 
capable of graphic arts. The single functions are regulated on the 
basis of three-front-sided arranged push buttons. 

The  menu  system  is  closed,  thereby  one  can  "browse"  forward 
as well as backward by the single setting menus to reach to the 
desired setting point. 

8.2 Structure of the menu system 

 

Fig. 3 Display and operating foil 

barograph 

display 

right button 

menu button 

left button 

Fig. 2 Battery case cap and communication interface 
Fig. 2 Battery case cap and communication interface

battery case 

battery case cap 

fixing screw 

connecting cable 

comm. - 

interface 

insulation foil 

protection screwing 

battery 

timer / battery 

pressure value 

time 

temperature 

Fig. 1 Lock against protection 

lock against 

protection 

pressure transmit-

ter module 

display 

PCE Americas Inc.

711 Commerce Way 

Suite 8 

Jupiter

FL-33458

USA

From outside US: +1

Tel: (561) 320-9162

Fax: (561) 320-9176

[email protected]

www.pce-instruments.com/english

www.pce-instruments.com

PCE Instruments UK Ltd.

Unit 11 

Southpoint Business Park 

Ensign way

Hampshire / Southampton

United Kingdom, SO31 4RF

From outside UK: +44

Tel: (0) 2380 98703 0

Fax: (0) 2380 98703 9

 [email protected]

Reviews: