A-type durometer applies to general rubber, synthetic rubber, soft rubber, 

poly-resin, leather, wax, etc. 

C-type  durometer  applies  to  rubber  and  contains  the  tiny  hole  material 

that that the vesicant is made in being used for plastic 

D-type durometer applies to general hard rubber, hard resin, acryl, glass, 

thermoplastics, printing plates, fibres, etc. 

1.Overview 

Shore  durometer  is  used  to  test  hardness  of  sulfureted  rubber  and 

plastics, such instrument has A-type/D-type/C-type. 

A-type and D-type apply to test the low and medium level hardness and 
high level hardness of materials. 
C-type  applies  to  test  the  hardness  of  shoe-making  porous  materials 
made by vesicants within plastics when the compression rate is 50% and 
stress is beyond 0.049MPa. 

2.Main parameters 

Dial value:1-100 degree 
Pointer journey:0-2.5mm 
Stress at the end of pointer:0.55N-8.06N of A-type and C-type;0-44.5N of 
D-type 

3.Operation instructions 

Put the sample on a solid plane,hold the durometer to make the pointer 

at  least  12mm  away  from  the  sample  edge,and  stably  put  the  legs  of 

durometer on the sample.Thus the pointer straightly goes into the sample till 

the legs completely get to the sample.Check the reading in 1s.Repeat shuch 

test 5 times at different place which are at least 6mm away from the test 

point and find the average reading(the distance between porous materials is 

at least 15mm). 

 
 

4.Notices 

Check to make sure before test that figure indicates zero(if not indicates 

zero,users may press “ ZERO” key).When press the durometer on the glass 
plane,the  figure  should  indicate  100  degree(the  top  of  pointer  an  its  legs 
should  tightly  touch  the  glass).If  it  does  not  indicates  ZERO  or  100 
degree,slightly  press  the  pinter  several  times;if  there  is  still  error,it  is 
advisable to send it back to the manufacturer. 

If  it  allows,rubber  samples  should  be  under  the  laboratory  standard 

temperature.If  the  test  result  of  the  sample  by  A-type  durometer  goes 
beyond 90 degrees,it is advisable to use D-type durometer.When the test of 
the sample by D-type durometer goes belw 20 degrees,it is advisable to use 
A-type durometer.If the figure on A-type duorometer is below 10 degrees,it is 
incorrect  and  can’t  be  used.When  all  the  tests  are over,users  should  clean 
the gauge,put it back into the box and keep it in a dry place. 

 

 

   
       

  

 

     