background image

 

 

©  PCE Instruments 

 

10  Data group  

With this coating thickness gauge, the measured values are saved automatically. Various data 
groups are available for this purpose. To make settings to the data memory, go to the menu under 
the item "Data Group". Here, you can make the following settings: 

Course (History) 

Here you can view the last measured values 
belonging to the set measuring group. 

Selected group (Select Grp XXX) 

Here you set the current measuring group. 
If you make settings in the menu item "Data 
group", these always refer to the data group 
selected here. The measured values are also 
saved in the data group selected here. 

Delete last value (Remove latest) 

With  this  function,  you  can  delete  the  last 
measured value. 

Delete group (Erase group) 

The entire measuring group is deleted here. 

Erase memory completely (Erase all) 

Here, you can clear the entire memory. 

11  Limit value alarm  

To set limit values, first go to the "limit alarm" menu. Under the item "Alarm on", you can activate 
the limit alarm function. You can set the limits as follows: 

Upper limit value (Up limit)  

Here, you can set the upper limit value. 

Lower limit value (Low limit)  

Here, you can set the lower limit value. 

If the measured value is within the limit range, the LED indicator  briefly flashes in green. If the 
measured value is outside the limit values, the LED indicator  briefly flashes in red. The display 
shows whether the current measured value is too high or too low.  

12  Convert unit  

To change the unit, go to the menu. Under the item "Unit", you can choose between the units µm 
and millimetre. 

13  Backlight 

The backlight switches on and off automatically. It is not possible to adjust this.  

14  Software  

The data from the coating thickness gauge can also be exported. For this purpose, there is the 
possibility of Bluetooth transfer to an Android or iOS device. It is also possible to transfer the data 
to a PC via the micro USB interface. 
 

 

Summary of Contents for PCE-CT 21BT

Page 1: ...Coating Thickness Gauge User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Las...

Page 2: ...off 4 7 Measurement 4 7 1 Continuous measurement 4 8 Menu 4 8 1 Menu structure 5 9 Calibration 5 9 1 Zero point calibration 6 9 2 One point calibration 6 9 3 Five point calibration 6 10 Data group 7...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...Fe Fe NFe Fe NFe Probe external internal external Measurement range 0 1500 m Resolution 0 1 m within meas range 0 99 9 m 1 m within meas range 100 1500 m Accuracy 1 m 2 of the coating thickness Units...

Page 5: ...terial m Unit Adjustable units m and mil v Limit value Is displayed when the set limit value is exceeded or fallen below 17 Number of measurements Number of measured values 4 1 Key description Symbol...

Page 6: ...ccur To carry out further measurements place the sensor on the next measuring spot Important In order to obtain the best results it is important to take measurements on a smooth and non slip surface D...

Page 7: ...OFF Unit m mil Device information About Sensor type Type Serial number Serial Hardware version Hardware Software version Software 9 Calibration To perform a calibration the meter has various function...

Page 8: ...asked whether you want to accept the calibration Press the key to save the change To discard the setting press the key 9 3 Five point calibration To perform a five point calibration first go to the ca...

Page 9: ...e memory 11 Limit value alarm To set limit values first go to the limit alarm menu Under the item Alarm on you can activate the limit alarm function You can set the limits as follows Upper limit value...

Page 10: ...this is not available the driver may be missing You can find the driver in the program folder under the driver folder If you have a 64 bit system use the installation file CP210xVCPInstaller_x64 If y...

Page 11: ...le device first download the Coatingmaster app Then switch on the Bluetooth function in the measuring device Now connect the coating thickness gauge via the app To do this select Coating_XXXX in the a...

Page 12: ...ngs to the app Setting Meaning Measuring sounds During a measurement a sound can be heard Measuring vibration When a measurement is made the vibration alarm is activated Alarm sounds When the limit is...

Page 13: ...any measurements at the edge Incorrect placement of the probe Always align the probe vertically and quickly with the surface Do not perform a measurement by force The probe must touch the test object...

Page 14: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 15: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Reviews: