Suggested  number  of  measurements  for  reinforced  and  non-reinforced  plastic  (According  to 
GB/T3854-2005) 
 
Non-reinforced 
plastic 
(hard plastic) 

Min. number of 
measurements 

Reinforced plastic 
(glass, steel)  

Min. number of 
measurements 

20 

30 

29 

30 

40 

22 

40 

50 

16 

50 

60 

10 

60 

70 

70 

 

 

80 

 

 

 

Calibration 

 
The calibration of this tester includes low end calibration, high end calibration, and display value 
calibration.  The  device  comes  factory  calibrated. If  the  reading  is  out  of  the  indicated  range  of 
the  test  blocks  when  checking  the  accuracy  or  if  you  replaced  the  pressure  pin  you  need  to 
calibrate the device. 
 

NOTE:  It  is  necessary  to  loosen  the  main  spindle  adjusting  screw  before  performing  a 
zero calibration or a full scale calibration. 

6.1 

Low end calibration 

Press and hold the Power / Menu key for about 9 seconds until 

‘’

’ is shown on the display. 

Release  the  key  and  a  digit  appears  on  the  display.  Press  the  Maximum  /  Plus  key  or  the 
Average  /  Minus  key  to  adjust  the  digit  until  it  equals  the  indicated  value  on  the  pin  length 
calibration block. Press the Power / Menu key to exit.  
Use  a  small  screw  driver  to  loosen  the  main  spindle  adjusting  screw  in  counter-clockwise 
direction until the end. Put the pin tube into the hole of the pin length calibration block and push 
the  tester  so  that  the  pin  draws  back  completely  into  the  pin  tube.  The  reading  value  should 
equal  the  thickness  value  indicated  on  the  pin length calibration  block. If it is  exceeded,  press 
the  low  end  calibration  button.  The  standard  value  will  be  displayed.  Release  the  low  end 
calibration button. The low end calibration is now completed. 

6.2 

High end calibration 

Use  a  small  screw  driver  to  loosen  the  main  spindle  adjusting  screw  in  counter-clockwise 
direction until the end. Place the tester on a solid flat surface, such as  a glass plate. Push the 
tester so that the pin draws back completely into the pin tube: 

‘100.0’ should be shown on the 

display.  If  it  is  exceeded,  press  the  high  end  calibration  button  and 

‘100.0’  will  be  displayed. 

Release the high end calibration button. The high end calibration is now completed. 

6.3 

Display value calibration 

Use  a  small  screw  driver  to  adjust  the  main  spindle  adjusting  screw.  The  display  value 
decreases when turning it clockwise and increases when turning it counter-clockwise. Adjust the 
screw until the measured value is in the indicated range of the 2 standard blocks. 
 

 

Summary of Contents for PCE-1000N

Page 1: ...N Barcol Hardness Tester User Manual User manuals in various languages Deutsch fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Last change 13...

Page 2: ...4 3 3 Display 5 4 Getting started 5 4 1 Power supply 5 4 2 Preparation 5 5 Operation 6 5 1 Measurement 6 5 2 Further Functions 6 6 Calibration 7 6 1 Low end calibration 7 6 2 High end calibration 7 6...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...ution 0 1 HBa Measurement error 81 88 HBa 1 HBa 42 48 Hba 2 HBa Repeatability error 81 88 HBa 1 5 HBa 42 48 HBa 2 5 Hba Operating conditions temperature humidity 0 50 C 50 RH Power supply 2 x 1 5 V AA...

Page 5: ...ead The effective measuring range is equivalent to a Brinell hardness of 25 150HB The max hold function shows the maximum value during the measurement The PCE 1000N can be used for measuring not only...

Page 6: ...0076 HBa Barcol hardness value H indentation depth mm 0 0076 indentation depth at 1 Barcol 2 1 housing screws 2 2 high end calibration button 2 3 low end calibration button 2 4 spring tube 2 5 lower s...

Page 7: ...lat surface and take a measurement The reading should be in the indicated range of two hardness blocks If it is out of the range a calibration according to chapter 6 is needed The sample surface shoul...

Page 8: ...asurements taken reaches the set value of the statistical measurement number N Now the device calculates the average value of the measurements taken The result is shown on the right side of the displa...

Page 9: ...se a small screw driver to loosen the main spindle adjusting screw in counter clockwise direction until the end Put the pin tube into the hole of the pin length calibration block and push the tester s...

Page 10: ...th a small screwdriver a total of 12 Take off the case and take out the spring tube Loosen the lower spindle connecting screw with a spanner and take out the lower spindle Now you can remove the used...

Page 11: ...version between different hardness systems is not possible Therefore the following conversion table is for reference only It is recommended to determine the actual hardness values by taking measuremen...

Page 12: ...6 92 67 57 62 10 6 69 68 93 68 60 65 11 0 71 70 94 69 62 67 11 4 73 72 95 70 64 70 11 8 17 75 74 97 71 67 72 12 2 23 76 75 98 72 69 75 12 6 28 78 77 99 73 72 78 12 9 33 80 79 100 74 75 81 13 3 38 81 8...

Page 13: ...PCE Instruments 11 93 18 4 103 100 94 18 6 104 101 95 18 7 105 102 96 18 9 106 102 97 19 0 106 103 98 19 2 107 99 19 3 107 100 19 4...

Page 14: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 15: ...4 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english www pce ins...

Reviews: