22
Characteristics ML 5032 C
USB
Measuring Quantity
Measuring Range
Measuring Range
Resolution
Resolution
in Lux (lx)
in footcandle (fc)
in lx
in fc
Illuminance
I
0.1... 199.9
0.01... 19.99
0.1
0.01
II
1... 1 999
0.1... 199.9
1
0.1
III
10... 19 990
1... 1 999
10
1
IV
100... 199 900
10... 19 990
100
10
in Candela/m² (cd/m²)
in footlambert (fL)
in cd/m²
in fL
Luminance
I
1... 1 999
0.1... 199.9
1
0.1
with luminance
II
10... 19 990
1... 1 999
10
1
attachment
III
100... 199 900
10... 19 990
100
10
for cd/m² or fL
IV
1000... 1 999 000
100... 199 900
1000
100
Most Important Error Limits ML 5032 C
USB
Characteristics
Admissible Error according to
DIN 5032 Klasse C
Typical Error
MAVOLUX 5032 C
USB
V(
)-Adaptation (
f
1
)
9%
7.5%
True Cosine Evaluation (
f
2
)
6%
2.0%
Linearity (
f
3
)
5%
1.5%
Adjustment Error (
f
11
)
2%
1.0%
Overall Error (
f
ges
)
20%
~ 15.0%