15
8.5.3
File processing
•
Press the Menu button to enter the Instrument Menu Settings screen.
•
Use the navigational buttons to navigate the menu to the data icon.
•
Press the menu button (
≡) to enter the data settings.
•
Select the File No. setting and
use the ◄ or ► buttons to change the file set number.
The device supports up to 5 file sets.
•
Select the File Browser and use the
▼ to ▲ buttons to navigate the 5 file sets. Use the
menu button (≡) to view a file set. Each file set can contain up to 28 data sets.
•
Select File Clear and use the menu button
(≡) to confirm the process. The entire
selected file set will be cleared.
9
Calibration
The device and impact probe must be calibrated with a coating thickness matrix before first use
or after a prolonged period of non-use.
•
Press the Menu button to enter the Instrument Menu Settings screen.
•
Use the navigational buttons to navigate the menu to the calibrate icon.
•
Press the menu button (
≡) to enter the data settings.
6 measurements must be performed vertically down into the provided substrate; the first is a zero
calibration (without the substrate).
•
Use the
▲ or ▼ buttons to adjust the value to the nominal thickness of the standard
test substrate for each measurement.
•
Perform a total of 5 measurements; one test for one test foil.
•
Press the menu button (
≡) to save and exit the calibration process.
10 Maintenance and Repair
10.1
Environmental requirements
Strictly avoid collisions, heavy dust, moisture, strong magnetic fields, oil stains, and other dirtying
or damaging factors.