background image

 

 

©  PCE Instruments 

 

On/off 

To turn on the meter, press the on/off key. To turn off the meter, press and hold the on/off key 
until 

“OFF“ is displayed. 

Automatic Power Off 

The gloss meter has an automatic power off function to save battery power. If no key is pressed 
for 10 minutes when a single measurement is made, the meter will turn off automatically. When 
a  continuous  measurement  is made,  the  meter  will  power  off  automatically  after  an  hour.  This 
function cannot be deactivated in measuring mode. 

Measurement 

The gloss meter hast wo measuring functions: 
In single measuring mode, the user can press the MEAS key to make a measurement. 

 will 

be displayed during a measurement. The reading will be saved directly after a measurement. 
In  continuous  measuring  mode,  the  meter  automatically  makes  continuous  measurements 
without  requiring  the  user  to  make  any  entries.  A  measurement  is  made  approx.  every  three 
seconds. The readings are saved continuously. 
To  switch  between  these  measuring  functions,  press  and  hold  the  on/off  key  until 

“SC”  is 

displayed after 

“OFF” was shown. 

To change the measuring angle, press the ANGLE key. 

Calibration 

The gloss meter should be calibrated regularly. To calibrate the meter, place it onto the calibration 
standard. Now make one measurement per angle and compare the readings to the values on the 
reference. 
If  one  reading  is  outside the  parameters, you can  automatically set  the value for  this  angle by 
pressing the CAL key while the meter is still on the reference. 

10  New calibration standard 

If  you  want  to  use  a  new  calibration  standard,  this  standard  must  first  be  added.  To  add  the 
standard, press and hold the on/off key (while the meter is on) until 

“CAL” appears on the display 

(after 

”OFF“ and ”SC“). Then press the CAL key. “CAL” will be displayed, followed by the angle 

and  the  pre-set value. You  can  now  change  the  reference value  with the  arrow  keys.  With  the 
ANGLE key, you can select the next measuring  angle. After setting the reference values for all 
measuring angles, press the ANGLE key to confirm the settings and return to measuring mode. 
 

 

Summary of Contents for GM 55

Page 1: ...Gloss Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 21...

Page 2: ...evice description 3 5 Display description 4 6 On off 5 7 Automatic Power Off 5 8 Measurement 5 9 Calibration 5 10 New calibration standard 5 11 Saving viewing deleting measured data 6 12 Recharging th...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...20 45 75 Measuring range 0 200 gloss units Resolution 0 1 gloss units Accuracy 1 gloss units on reference standard Repeatability 0 5 gloss units 0 99 9 GE Measuring surface 7 x 14 mm elliptical Data...

Page 5: ...to set measuring angle ANGLE 4 CAL key for calibration and to decrease parameters 5 MEAS key to make a measurement 6 DEL key to delete reading 7 RD key to view memory and increase parameter 8 USB inte...

Page 6: ...tion 1 Indicates that you are in calibration mode 2 Reading 3 Measuring mode 4 Measuring angle 5 Continuous mode 6 Memory readout mode 7 Memory item no 8 Indicates that battery must be recharged 9 Ind...

Page 7: ...ed continuously To switch between these measuring functions press and hold the on off key until SC is displayed after OFF was shown To change the measuring angle press the ANGLE key 9 Calibration The...

Page 8: ...ode Note If you try to enter memory readout mode when no measuring data are available the display will show Err1 Memory readout mode can only be opened if data have been saved 12 Recharging the batter...

Page 9: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 10: ...583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info pce instruments co uk info pce italia it www pce instruments com deutsch www pce instrumen...

Reviews: