Drawing Number: 21110 
Revision: B 

 

OPERATION MANUAL FOR 

QUARTZ PRESSURE SENSOR 

Series 118 & 119

 

 

 

3.2 

PREPARING NEW MOUNTING  

 

PORTS

 

 

Refer  to  the  installation  drawing  provided  in  this 
manual for instructions on mounting hole preparation.  
For  best  results,  do  not  deviate  from  the  steps 
outlined in this drawing. 

 

To assist new mounting port preparation, PCB offers 
the  040B20  Tooling  Kit  (040B21  for  metric mount).  
These  installation  kits  provide  all  of  the  necessary 
tooling  required  to  drill,  ream,  and  tap  the  mounting 
ports for proper installation of the 108, 109, 118 and 
119 Series transducers.   

 

Use  good  machining  practice  in  preparation  of  the 
mounting  port,  paying  particular  attention  to the seal 
surface.    It  is  important  that this surface be perfectly 
smooth  and  free  from  tool  chatter  marks,  nicks  and 
other  imperfections  which  might  cause  leaks  at  high 
pressures. 
 

3.3 

RECESSED MOUNT

 

 

The  recessed  installation  is  shown  in  the  Recessed 
Mount  figure.    This  type  of  mounting  protects  the 
sensor  diaphragm  from  the  effects  of  high-flash 
temperatures  and  particle  impingement  due  to  blast 
effects, thereby prolonging sensor life. 

 

The  recommended  range  of  passage  diameters  is 
0.090 to 0.125 inch (2.29 to 3.18 mm) diameter. 

 

The  limitation  in  this  type  of  installation  lies  in  the 
frequency-limiting  effects  of  the  passage  due  to  its 
length. 

 

The  passage  behaves  like  an  underdamped  second 
order 

system; 

the  resonant  frequency  being 

determined  by  the  passage  length.    The  length  may 
have a limiting effect on pressure pulse rise time and 
cause  passage  ringing  in  cases  where  the  passage  is 
too long. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Recessed Mount 

 
The following relationship approximates this resonant 
frequency (f

 r

): 

 

(Hz)

L

4

V

f

r

 

(EQ. 1) 

 
Where:  f

 r

 = Resonant frequency of passage (Hz) 

V = Velocity of sound in air (ft/sec) 

L = Length of column (ft) 

For air at room temperature, (EQ. 1) becomes: 
 

L

00

3

3

f

r

 

 

(EQ. 2) 

 
Where:   L = Passage length (in) 

0.010 

MIN 

Summary of Contents for UP118B11

Page 1: ...or 80k psi 0 11 pC psi floating Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001...

Page 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...hese sensors a high natural frequency and linearity 3 0 INSTALLATION With a 3 8 24 M10x1 0 for metric mount mounting thread and flush diaphragm design the 118 119 Series transducers mount directly in...

Page 7: ...aks at high pressures 3 3 RECESSED MOUNT The recessed installation is shown in the Recessed Mount figure This type of mounting protects the sensor diaphragm from the effects of high flash temperatures...

Page 8: ...emperature effects are usually longer term and will show up as a baseline shift long after the event to be measured has passed For flush mount installations a silicone rubber coating approximately 0 0...

Page 9: ...ARITY Since most charge amplifiers are inverting the input signal 118 119 Series are designed to produce a negative output voltage for increasing pressure at the diaphragm Special positive output vers...

Page 10: ......

Page 11: ......

Page 12: ......

Reviews: