PCB Piezotronics UM165A01 Installation And Operating Manual Download Page 1

Model UM165A01

CHARGE PRESSURE SENSOR

Installation and Operating Manual

For assistance with the operation of this product,

contact PCB Piezotronics, Inc.

Toll-free: 800-828-8840

24-hour SensorLine: 716-684-0001

Fax: 716-684-0987

E-mail: [email protected]

Web: www.pcb.com

Summary of Contents for UM165A01

Page 1: ...E SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Page 2: ...repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps...

Page 3: ...ry customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...rs The short rigid quartz element and stiff diaphragm give the unit a 250 kHz natural frequency and good linearity even below 1 000 psi 3 0 INSTALLATION When choosing an installation method the advant...

Page 7: ...it free from tool chatter marks NOTE It is important that this surface be perfectly smooth and free from nicks or other discontinuities that could cause leaks at high pressures 3 3 Recess Mount The r...

Page 8: ...be thermally insulated with DC 4 silicone grease or its equivalent In place of the silicone rubber one layer of black vinyl electrical tape on the diaphragm may provide sufficient insulation 3 6 Cable...

Page 9: ...ess the ground button of the charge amplifier and adjust electrical zero if necessary Range the amplifier to give the necessary full scale voltage For normal drift free operation switch the charge amp...