background image

Model S102A07

High resolution ICP® pressure sensor, 50 psi, 100 mV/psi, 3/8-24 mtg thd

Installation and Operating Manual

For assistance with the operation of this product,

contact PCB Piezotronics, Inc.

Toll-free: 800-828-8840

24-hour SensorLine: 716-684-0001

Fax: 716-684-0987

E-mail: [email protected]

Web: www.pcb.com

Summary of Contents for S102A07

Page 1: ...psi 100 mV psi 3 8 24 mtg thd Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...t Preload Sleeve Typical ICP Probe Style Sensor Each model utilizes the basic ICP pressure probe as shown in above figure The pressure probe consists of the Model 112A high sensitivity acceleration co...

Page 7: ...s the end of the diaphragm without affecting response or sensitivity A silicone rubber coating approximately 010 thick has also been proven effective in many applications General Electric RTV type 106...

Page 8: ...and 2 If AC coupled at power unit the coupling time constant Consult Section 7 0 in Guide G 0001B for detailed explanation of low frequency characteristics of ICP instruments 7 0 CALIBRATION Piezoele...

Page 9: ...ENSORS Models 102A02 A05 A07 A09 Models 112A21 A22 A23 4 8 0 MAINTENANCE The miniature size sealed construction precludes field maintenance Should service be required return unit to factory with note...

Page 10: ...4LStainless Steel ThreadAdaptor Material 17 4 Stainless Steel 17 4 Stainless Steel Diaphragm Invar Invar Sealing Hermetic Hermetic Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 3...

Page 11: ......