Model Number

480E09

BATTERY-POWERED SIGNAL CONDITIONER

Revision: V
ECN #: 45339

Performance

ENGLISH

SI 

Channels

1

1

Frequency Range(-5 %)(x1, x10 Gain)

0.15 to 100,000 Hz

0.15 to 100,000 Hz

[6]

Frequency Range

(-10 %)(x100 Gain)

0.15 to 50,000 Hz

0.15 to 50,000 Hz

[7]

Voltage Gain(± 2 %)

1:1

1:1

Voltage Gain

(± 2 %)

1:10

1:10

Voltage Gain

(± 2 %)

1:100

1:100

Fault/Bias Monitor/Meter(± 1 V)(midscale)

13 VDC

13 VDC

Environmental

Temperature Range

32 to 120 °F

0 to 50 °C

Electrical

Excitation Voltage(To Sensor)

25 to 29 VDC

25 to 29 VDC

[1]

Constant Current Excitation(To Sensor)

2.0 to 3.2 mA

2.0 to 3.2 mA

[2]

Discharge Time Constant

>7 sec

>7 sec

[3][4]

DC Offset(Maximum)

<30 mV

<30 mV

[3]

Spectral Noise(1 Hz)(Gain 1)

.25 µV/

Hz

-132 dB

Spectral Noise

(10 Hz)(Gain 1)

.07 µV/

Hz

-143 dB

Spectral Noise

(100 Hz)(Gain 1)

.05 µV/

Hz

-146 dB

Spectral Noise

(1 kHz)(Gain 1)

.04 µV/

Hz

-148 dB

Spectral Noise

(10 kHz)(Gain 1)

.03 µV/

Hz

-150 dB

Broadband Electrical Noise(1 to 10,000 Hz)(Gain x1)

3.25 µV rms

-110 dB/rms

Spectral Noise(1 Hz)(Gain 10)

2.2 µV/

Hz

-113 dB

Spectral Noise

(10 Hz)(Gain 10)

2.0 µV/

Hz

-114 dB

Spectral Noise

(100 Hz)(Gain 10)

1.1 µV/

Hz

-119 dB

Spectral Noise

(1 kHz)(Gain 10)

.55 µV/

Hz

-125 dB

Spectral Noise

(10 kHz)(Gain 10)

.3 µV/

Hz

-130 dB

Broadband Electrical Noise(1 to 10,000 Hz)(Gain x10)

49 µV/rms

-86 dB/rms

Spectral Noise(1 Hz)(Gain 100)

20 µV/

Hz

-94 dB

Spectral Noise

(10 Hz)(Gain 100)

19 µV/

Hz

-94 dB

Spectral Noise

(100 Hz)(Gain 100)

12 µV/

Hz

-98 dB

Spectral Noise

(1 kHz)(Gain 100)

5.5 µV/

Hz

-105 dB

Spectral Noise

(10 kHz)(Gain 100)

2 µV/

Hz

-114 dB

Broadband Electrical Noise(1 to 10,000 Hz)(Gain x100)

569 µV/rms

-65 dB/rms

Power Required(Standard)

Internal Battery

Internal Battery

Internal Battery(Type)

9V

9V

Battery Life(Standard Alkaline)

50 hours

50 hours

Power Required(Alternate)

DC power

DC power

DC Power

15 mA

15 mA

[5]

Internal Battery(Quantity)

3

3

DC Power

18 to 30 VDC

18 to 30 VDC

[5][1]

Physical

Electrical Connector(Input, sensor)

BNC Jack

BNC Jack

Electrical Connector

(Output, scope)

BNC Jack

BNC Jack

Electrical Connector

(External Power, DC)

3.5 mm Diameter Miniature 

Jack

3.5 mm Diameter Miniature 

Jack

Electrical Connector

(Battery Charger)

#722 Switchcraft Jack

#722 Switchcraft Jack

Size (Depth x Height x Width)

2.4 in x 4.0 in x 2.9 in

6.1 cm x 10 cm x 7.4 cm

Weight(Including Batteries)

0.7 lb

0.3 kg

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

 is a registered trademark of PCB Group, Inc. 

NOTES:

[1]Excitation voltage to sensor limited by optional DC power voltage.
[2]Through internal current limiting regulator.
[3]With 1M ohm load.
[4]Un-buffered output, read out device input impedance affects discharge time constant and low 

frequency response of unit.

[5]Provided by optional external DC power supply.
[6]Low frequency response specified into 1M ohm load.
[7]After Serial Number 24699, previously HFR was 100kHz.
[8]See PCB Declaration of Conformance PS024 for details.

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

[8]

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard 

model except where noted below. More than one option may be used.

Entered: LK

Engineer: CPH

Sales: ML

Approved: JWH

Spec Number:

Date: 5/23/2016

Date: 5/23/2016

Date: 5/23/2016

Date: 5/23/2016

480-5090-80 

Summary of Contents for R480E09

Page 1: ...signal conditioner gain x1 x10 Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...ICP Signal Conditioners Operating Manual with Enclosed Warranty Information 3425 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 828 8840 MANUAL NUMBE...

Page 7: ...tents Introduction page 3 Description page 4 Operation page 5 Performance Limits page 8 Battery Considerations page 8 Optional Power Connections page 9 Warranty page 10 Calibration Service page 10 Ret...

Page 8: ...They are powered internally by three replaceable batteries and provide constant current excitation to the charge amplifier within ICP sensors or to in line and adaptor style amplifiers and decouples...

Page 9: ...el meter indicates the voltage output when connected to a sensor and is color coded for various fault conditions shown in Figure 1 2 Optional connections include sockets for EXT PWR and BATT CHG A cen...

Page 10: ...age with full scale set at 27 volts 27 volts for new batteries see Figure 1 When an ICP sensor is connector to the input XDCR jack the meter will indicate approximately mid scale 11 volts nominal if t...

Page 11: ...itor is fully charged Such drifting is considered normal operation Coupling Time Constant TC AC Coupled The coupling discharge time constant DTC is the product of the capacitor 47 F and the parallel i...

Page 12: ...e 5 the low frequency response of the coupling circuit is determined by the relationship in Equation 2 This requires use of a BNC Tee Figure 6 below provides a comparison of AC vs DC coupling the same...

Page 13: ...driven but it sacrifices some high frequency response PCB line powered signal conditioners 482 Series 483 Series should be considered for cable lengths beyond 100 feet 31 meters as they provide up to...

Page 14: ...not ship with rechargeable batteries and PCB external battery chargers are suitable for use with Nickel Cadmium Ni Cad rechargeable batteries only It should not be used with alkaline or other non rec...

Page 15: ...lt a factory PCB applications engineer concerning the situation as certain problems can often be corrected with simple on site procedures PCB will perform calibrations on model 480C02 on request Typic...

Page 16: ...funded PCB offers to all customers at no charge 24 hour phone support This service makes product or application support available to our customers day or night seven days a week When unforeseen proble...

Page 17: ...rnal Battery Internal Battery Internal Battery Type 9V 9V Battery Life Standard Alkaline 50 hours 50 hours Power Required Alternate DC power DC power DC Power 15 mA 15 mA 5 Internal Battery Quantity 3...

Page 18: ......

Reviews: